Inhibitory effects of a minimally invasive implanted tibial nerve stimulation device on non-nociceptive bladder reflexes in cats
Related Experiment Video
Updated: Dec 7, 2025

Urinary Bladder Distention Evoked Visceromotor Responses as a Model for Bladder Pain in Mice
Published on: April 27, 2014
Xiaoping Wan1,2,3,4, Yanan Liang1,2,3,4, Xing Li5,6,7,8
1Department of Urology, China Rehabilitation Research Center, Rehabilitation School of Capital Medical University, Beijing, 100068, China.
Objective:
The present study investigated the inhibitory effects of a novel minimally invasive implanted tibial nerve stimulation device on non-nociceptive bladder reflexes in cats.
Materials And Methods:
A wireless minimally invasive implanted nerve stimulator was implanted adjacent to the left tibial nerve in seven cats. Multiple cystometrograms (CMGs) were obtained to determine the inhibitory effects of tibial nerve stimulation (TNS) at different frequencies and intensities on the micturition reflex by infusing normal saline (NS).
Results:
TNS at 6 Hz did not significantly change the bladder capacity (BC) compared to the control level at the intensity threshold (T), while TNS significantly (P < 0.05) increased the BC to 158.89 ± 20.57% of the control level at 2 T. When stimulated at 15 Hz, TNS did not significantly (P > 0.05) change the BCs at 1 T and 2 T.
Conclusions:
The minimally invasive implanted TNS device was shown to be effective in inhibiting the micturition reflex under physiologic conditions. Further studies are warranted to determine the inhibitory effects of TNS on nociceptive bladder reflexes.