Distortion-free 3D diffusion imaging of the prostate using a multishot diffusion-prepared phase-cycled acquisition
Elisa Roccia1, Radhouene Neji1,2, Thomas Benkert3
1School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom.
Magnetic Resonance in Medicine
|September 29, 2020
Summary
This study introduces a new method for creating distortion-free 3D apparent diffusion coefficient (ADC) maps of the prostate using a multishot diffusion prepared-gradient echo (msDP-GRE) sequence and ADC dictionary matching, offering improved imaging accuracy.
Area of Science:
- Magnetic Resonance Imaging
- Medical Physics
- Radiology
Background:
- Prostate imaging requires accurate apparent diffusion coefficient (ADC) maps.
- Geometric distortion in diffusion-weighted imaging (DWI) can affect diagnostic accuracy.
- Existing methods may be limited by distortion and acquisition time.
Purpose of the Study:
- To develop and validate a novel technique for generating 3D distortion-free ADC maps of the prostate.
- To utilize a multishot diffusion prepared-gradient echo (msDP-GRE) sequence combined with ADC dictionary matching.
- To overcome limitations of current prostate diffusion imaging.
Main Methods:
- Implemented a 3D Cartesian, centric k-space trajectory with center oversampling in the msDP-GRE sequence.
- Employed k-space center averaging and phase cycling to mitigate motion and eddy current artifacts.
- Used Extended-Phase-Graph (EPG) simulations and ADC dictionary matching for T1 effect compensation and undersampled reconstruction.
- Validated the approach in a kiwifruit phantom and tested in vivo feasibility in healthy subjects.
Main Results:
- The msDP-GRE sequence with dictionary matching successfully recovered signal magnitude and produced distortion-free prostate ADC maps.
- No significant difference was found between ADC values from msDP-GRE and clinical standard single-shot echo-planar imaging (ssDW-EPI).
- In vivo scans showed no geometric distortion, and prostate ADC values aligned with published literature.
Conclusions:
- Achieved non-distorted 3D apparent diffusion coefficient (ADC) maps of the prostate.
- The multishot diffusion prepared (msDP) sequence with dictionary matching is effective for prostate imaging.
- This technique offers a promising alternative for accurate prostate diffusion MRI.