Comparison of anorectal function measured using wearable digital manometry and a high resolution manometry system
Ali Attari1, William D Chey2, Jason R Baker2
1Department of Mechanical Engineering, University of Michigan, Ann Arbor, MI, United States of America.
A new, low-cost digital manometry system effectively assesses anorectal function. This finger-based device offers comparable results to high-resolution systems for diagnosing constipation and incontinence with improved patient comfort.
Area of Science:
- Gastroenterology and Urology
- Medical Device Technology
Background:
- Assessing anorectal function is crucial for diagnosing conditions like chronic constipation and fecal incontinence.
- Existing high-resolution manometry systems are costly and may be inaccessible in primary care settings.
Purpose of the Study:
- To develop and validate a lower-cost, index finger-based "digital manometry" system for anorectal function assessment.
- To compare the efficacy and patient comfort of digital manometry against a high-resolution manometry system.
Main Methods:
- Digital manometry was developed using an index finger-based approach.
- Anorectal pressures and puborectalis myoelectric activity were measured in 16 participants.
- Data was compared with a 23-channel high-resolution anorectal manometry system using Bland-Altman analysis.
Main Results:
- Digital manometry showed acceptable limits of agreement with high-resolution manometry for resting, squeeze, and defecation maneuvers.
- Puborectalis myoelectric activity changes correlated with anal sphincter pressure increases during maximum squeeze.
- The digital system demonstrated an order of magnitude lower cost and similar patient comfort.
Conclusions:
- Digital manometry is a simple, inexpensive, point-of-service tool for evaluating anorectal function.
- It offers a viable alternative for assessing patients with chronic constipation and fecal incontinence.
- This technology can improve accessibility to manometry in diverse clinical settings.
More Related Videos
03:49Anogenital Distance and Perineal Measurements of the Pelvic Organ Prolapse POP Quantification System
Published on: September 20, 2018
10:07Cystometric and External Urethral Sphincter Measurements in Awake Rats with Implanted Catheter and Electrodes Allowing for Repeated Measurements
Published on: January 30, 2018
Related Concept Videos
Assessment of the Rectum and Anus
Rectal Inspection
Begin by inspecting the perianal and anal areas for color, texture, rashes,...
Urodynamic Studies: Uroflowmetry
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...