Does reducing radiation levels for procedures affect image quality and radiation to proceduralists? A double-blinded
Z Zhang1, C C Phang2, R Y Tan2
1Duke-NUS Medical School, National University of Singapore, 169857, Singapore.
Aim:
To evaluate the ultra-lose dose imaging protocol (ULDP), compared to the standard low-dose imaging protocol (LDP), which are used for haemodialysis access, in terms of radiation exposure and image quality.
Material And Methods:
This was a single-centre, institutional review board-approved, prospective, double-blinded randomised controlled study to compare radiation exposure and image quality of the ULDP and LDP. Ten proceduralists, two radiographers, and 11 nurses were enrolled. Radiation exposure during 80 procedures (40 angioplasties and 40 thrombolysis) was recorded (direct radiation to patients from protocol report and scattered radiation to participants from the RaySafe i2 real-time dosimetry system). Baseline characteristics of procedure were recorded. Image quality was assessed subjectively using questionnaires based on the five-point Likert scale after each procedure.
Results:
Compared with LDP, the use of ULDP was associated with a significantly lower rate of radiation exposure to proceduralists, patients, and scrub nurses (0.506±0.430 versus 0.847±0.965 μSv/s, p=0.044; 0.571±1.284 versus 1.284±1.007 mGy/s, p<0.001; and 0.052±0.071 versus 0.141±0.185 μSv/s, p=0.005, respectively). No significant difference in image quality or duration of procedure was observed (all p values >0.05).
Conclusion:
Compared with LDP, the use of ULDP was associated with a significantly lower rate of radiation exposure to proceduralists, patients, and scrub nurses without compromising the image quality or duration of procedure.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Blinding
Blind Procedures
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT