Assessment of a standalone photoplethysmography (PPG) algorithm for detection of atrial fibrillation on
J L Selder1, T Proesmans2, L Breukel3
1Amsterdam UMC, Vrije Universiteit, Department of Cardiology, Amsterdam, Netherlands.
Introduction:
Atrial fibrillation (AF) is the most common cardiac arrhythmia in the developed world. Using photoplethysmography (PPG) and software algorithms, AF can be detected with high accuracy using smartphone camera-derived data. However, reports of diagnostic accuracy of standalone algorithms using wristband-derived PPG data are sparse, while this provides a means to perform long-term AF screening and monitoring. This study evaluated the diagnostic accuracy of a well-known standalone algorithm using wristband-derived PPG data.
Materials And Methods:
Subjects recruited from a community senior care organization were instructed to wear the Wavelet PPG wristband on one arm and the Alivecor KardiaBand one-lead-ECG wristband on the other. Three consecutive measurements (duration per measurement: 60 s for PPG and 30 s for one-lead ECG) were performed with both devices, simultaneously. The PPG data were analyzed by the Fibricheck standalone algorithm and the ECG data by the Kardia algorithm. The results were compared to a reference standard (interpretation of the one-lead ECG by two independent cardiologists).
Results:
A total of 180 PPGs and one-lead ECGs were recorded in 60 subjects, with a mean age of 70±17. AF was identified in 6 (10%) of the users, two users (3%) were not classifiable by the PPG algorithm and 1 user (2%) was not classifiable by the one-lead ECG algorithm. The diagnostic performance (sensitivity/specificity/positive predictive value/negative predictive value/accuracy) on user level was 100/96/75/100/97% for the PPG wristband and 100/98/86/100/98% for the one-lead ECG wristband.
Conclusions:
In a small real-world cohort of elderly people, the standalone Fibricheck AF algorithm can accurately detect AF using Wavelet wristband-derived PPG data. Results are comparable to the Alivecor Kardia one-lead ECG device, with an acceptable unclassifiable/bad quality rate. This opens the door for long-term AF screening and monitoring.
More Related Videos
09:17High-Resolution Endocardial and Epicardial Optical Mapping in a Sheep Model of Stretch-Induced Atrial Fibrillation
Published on: July 29, 2011
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...