Finite Element Analysis of Insertion Angle of Absorbable Screws for the Fixation of Radial Head Fractures
Guang-Ming Xu1, Zi-Yang Liang2, Wei Li1
1Department of Orthopaedics, Shenzhen Pingle Orthopedic Hospital & Shenzhen Pingshan Traditional Chinese Medicine Hospital, Shenzhen, China.
Objective:
To investigate the biomechanical effects of different insertion angles of absorbable screws for the fixation of radial head fractures.
Methods:
The finite element models used to simulate the fractures were created based on CT scans. Two absorbable screws were used to fix and maintain the stability of the fracture, and the angles between the screws were set to 0°, 15°, 30°, 45°, 60°, 75°, and 90°. A downward force of 100 N was applied at the stress point, which was coupled with the surface, and the distal radius was limited to six degrees of freedom. The direction and location of the applied force were the same in each model. The values of the von Mises stress and peak displacements were calculated.
Results:
Under the applied load and different screw angles, the maximum von Mises stress in the screws was concentrated on the surface contacting the fracture surfaces. The maximum von Mises equivalent stress in the screw decreased when the angle increased from 0° (19.54 MPa) to 45° (13.11 MPa) and increased when the angle further increased to 90° (24.63 MPa). The peak displacement decreased as the angle increased from 0° (0.19 mm) to 45° (0.15 mm) and increased when the angle further increased to 90° (0.25 mm).
Conclusion:
The computational stress distribution showed that fixation with absorbable screws is safe for patients. Moreover, the minimum von Mises stress and displacements were generated when the angle between the screws was 45°; hence, this setting should be recommended for Mason type II radial fractures.
More Related Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Related Concept Videos
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Angle of Twist - Elastic Range
Screw: Problem Solving
To evaluate the...
Unsymmetric Bending - Angle of Neutral Axis
When a bending moment is applied at an angle θ concerning the vertical axis of a symmetrical member, it can be resolved into components along the member's principal...