Estimated Risk for Insulin Dose Error Among Hospital Patients Due to Glucose Meter Hematocrit Bias in 2020
Mark Inman1, Andrew W Lyon2, Oliver A S Lyon3
1From the Department of Pediatrics, University of Saskatchewan, Saskatoon, Saskatchewan, Canada (Inman).
Context.—:
Glycemic control requires accurate blood glucose testing. The extent of hematocrit interference is difficult to assess to assure quality patient care.
Objective.—:
To predict the effect of patient hematocrit on the performance of a glucose meter and its corresponding impact on insulin-dosing error.
Design.—:
Multilevel mixed regression was conducted to assess the extent that patient hematocrit influences Roche Accu-Chek Inform II glucose meters, using the Radiometer ABL 837 as a reference method collected during validation of 35 new meters. Regression coefficients of fixed effects for reference glucose, hematocrit, an interaction term, and random error were applied to 4 months of patient reference method results extracted from the laboratory information system. A hospital inpatient insulin dose algorithm was used to determine the frequency of insulin dose error between reference glucose and meter glucose results.
Results.—:
Fixed effects regression for method and hematocrit predicted biases to glucose meter results that met the "95% within ±12%" for the US Food and Drug Administration goal, but combinations of fixed and random effects exceeded that target in emergency and hospital inpatient units. Insulin dose errors were predicted from the meter results. Twenty-eight percent of intensive care unit, 20.8% of hospital inpatient, and 17.7% of emergency department results were predicted to trigger a ±1 insulin dose error by fixed and random effects.
Conclusions.—:
The current extent of hematocrit interference on glucose meter performance is anticipated to cause insulin error by 1-dose category, which is likely associated with low patient risk.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Hypoglycemia and Glucagon
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5%...
Bias in Epidemiological Studies