Molecular Beam Mass Spectrometry With Tunable Vacuum Ultraviolet VUV Synchrotron Radiation
Published on: October 30, 2012
A high-pressure reactor coupled to synchrotron radiation photoionization mass spectrometry
Rui You1, Shengsheng Yu2, Jiuzhong Yang2
1Hefei National Laboratory for Physical Sciences at the Microscale, Key Laboratory of Surface and Interface Chemistry and Energy Catalysis of Anhui Higher Education Institutes, CAS Key Laboratory of Materials for Energy Conversion and Department of Chemical Physics, University of Science and Technology of China, Hefei 230026, China.
Abstract:
A high-pressure reactor was designed and coupled to synchrotron radiation photoionization mass spectrometry (SR-PIMS), which realizes the molecular-beam sampling and detection of gaseous products of high-pressure reactions. The reaction pressure can be controlled by varying the size of the pinhole of the pressure-bearing pipe. As tested by the Fischer-Tropsch synthesis (FTS) catalyzed by Co/SiO2 at 230 °C, the reaction pressure of our setup can reach 1.3 MPa with a pinhole size of 50 µm and 0.16 MPa with a pinhole size of 150 µm. The FTS products were successfully online detected by SR-PIMS, and the photoionization efficiency spectra of selected products were acquired for unambiguous identification of the detected signals. Meanwhile, time-resolved SR-PIMS spectra were acquired with a temporal resolution of 10 s. The characterization results demonstrate that the product distribution (C2-C4, C5-C11, and C12+) of FTS depends on the reaction pressure, where a high pressure facilitates the formation of long-chain hydrocarbons. With the advantages of detecting unstable intermediates and distinguishing isomers, this setup will be useful for fundamental studies of high-pressure heterogeneous catalytic reactions.
More Related Videos
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Mass Spectrometers
Chemical Ionization (CI) Mass Spectrometry
High-Resolution Mass Spectrometry (HRMS)
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...