Active Endoscope Preserving Image Orientation for Endonasal Skull Base Surgery.
Summary
This study introduces an active endoscope for skull base surgery, improving visualization and safety. The new device offers a variable view and safer insertion, overcoming limitations of traditional rigid endoscopes.
Area of Science:
- Neurosurgery
- Medical Robotics
- Surgical Instrumentation
Background:
- Conventional endoscopes in endonasal skull base surgery have limitations, including a fixed field of view and potential invasiveness during insertion.
- These limitations can complicate procedures and increase the risk of damaging critical tissues.
Purpose of the Study:
- To propose and design an active endoscope mechanism to address the challenges of conventional endoscopes in endonasal skull base surgery.
- To enhance surgical precision, safety, and maneuverability within confined surgical spaces.
Main Methods:
- Development of a 4 mm diameter active endoscope with a 2-degree-of-freedom (DOF) tilting and rotation mechanism.
- Implementation of kinematics control for intuitive operation and a fixed image sensor to maintain orientation relative to gravity.
- Design allows for directional changes within a 6.5 mm radius.
Main Results:
- The active endoscope provides a variable direction of view, covering wider ranges and enabling a forward-looking perspective during insertion.
- Intuitive control was demonstrated through experiments using a master device, allowing users to observe targets without needing to swap the endoscope.
- The device's small form factor and maneuverability are suitable for use in narrow cavities alongside other surgical tools.
Conclusions:
- The developed active endoscope mechanism offers significant advantages over traditional rigid endoscopes for endonasal skull base surgery.
- It enhances safety, improves visualization, and increases maneuverability, potentially leading to better surgical outcomes.