Related Experiment Video
Updated: Dec 6, 2025

Ferromagnetic Bare Metal Stent for Endothelial Cell Capture and Retention
Published on: September 18, 2015
Finite element simulation and testing of cobalt-chromium stent: a parametric study on radial strength, recoil,
Avinash Kumar1, Naresh Bhatnagar1
1Department of Mechanical Engineering, Indian Institute of Technology-Delhi, New Delhi, India.
Abstract:
The effectiveness of cardiovascular stenting procedure depends on the crimping and expansion characteristics of a stent, influenced by its design parameters. In this study, CoCr stents are fabricated, crimped on a tri-folded balloon, and expanded using manual inflation device. Similarly, in the finite element model, a tri-folded balloon is used to expand the stent. The length and diameter are measured to evaluate the radial strength, recoil, foreshortening, and dogboning. The simulation and experimental results match satisfactorily. The validated FE model can be used with confidence to optimize future stent designs, thus reducing the number of testing and product development time.
More Related Videos
11:00Experimental Investigation of Secondary Flow Structures Downstream of a Model Type IV Stent Failure in a 180° Curved Artery Test Section
Published on: July 19, 2016
06:34Finite Element Modeling for the Simulation of the Quasi-Static Compression of Corrugated Tapered Tubes
Published on: January 6, 2023