Related Experiment Video
Updated: Dec 6, 2025

A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
Variability in haemoglobin concentration by measurement tool and blood source: an analysis from seven countries
Aviva I Rappaport1,2, Crystal D Karakochuk3, Sonja Y Hess4
1Strengthening Partnerships, Results, and Innovations in Nutrition Globally (SPRING), Arlington, Virginia, USA.
Objective:
We explore factors such as the blood sampling site (capillary vs venous), the equipment (HemoCue vs automated haematology analyser) and the model of the HemoCue device (201+ vs 301) that may impact haemoglobin measurements in capillary and venous blood.
Methods:
Eleven studies were identified, and bias, concordance and measures of diagnostic performance were assessed within each study.
Findings:
Our analysis included 11 studies from seven countries (Cambodia, India, The Gambia, Ghana, Laos, Rwanda and USA). Samples came from children, men, non-pregnant women and pregnant women. Mean bias ranged from -8.7 to 2.5 g/L in Cambodian women, 6.2 g/L in Laotian children, 2.4 g/L in Ghanaian women, 0.8 g/L in Gambian children 6-23 months and 1.4 g/L in Rwandan children 6-59 months when comparing capillary blood on a HemoCue to venous blood on a haematology analyser. Bias was 8.3 g/L in Indian non-pregnant women and 2.6 g/L in Laotian children and women and 1.5 g/L in the US population when comparing capillary to venous blood using a HemoCue. For venous blood measured on the HemoCue compared with the automated haematology analyser, bias was 5.3 g/L in Gambian pregnant women 18-45 years and 11.3 g/L in Laotian children 6-59 months.
Conclusion:
Our analysis found large variability in haemoglobin concentration measured on capillary or venous blood and using HemoCue Hb 201+ or Hb 301 or automated haematology analyser. We cannot ascertain whether the variation is due to differences in the equipment, differences in capillary and venous blood, or factors affecting blood collection techniques.
Related Concept Videos
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...
Hemoglobin
When all four heme groups are bound to oxygen, the resulting molecule is called oxyhemoglobin. As a result, arterial blood...
Factors Affecting Erythropoiesis
Several factors influence the erythrocyte production rate, with tissue oxygen level being among the most critical. Intense exercise or high altitudes can cause tissue hypoxia, which triggers the kidneys to release more erythropoietin (EPO) into the bloodstream.
EPO then...
Multiple Allele Traits
Blood Typing
Antigens are protein molecules that reside on the surface of red blood cells (RBCs). The ABO and Rh blood typing systems target...
Blood Transfusion and Agglutination
History
The history of blood transfusion dates back to the 17th century, when early attempts were made in animals. In 1818 James Blundell, a British doctor, performed the first successful human blood transfusion. Later in 1900, Karl...