Reliability of Smartphone Pulse Oximetry in Subjects at Risk for Hypoxemia
Aashna M Modi1, Renee D Kiourkas1, Jie Li2
1Respiratory Care Services, Rush University Medical Center, Chicago, Illinois.
Background:
Pulse oximeters are used to measure [Formula: see text] and pulse rate. These devices are either standalone machines or integrated into physiologic monitoring systems. Some smartphones now have pulse oximetry capabilities. Because it is possible that some patients might utilize this technology, we sought to assess the accuracy and usability of smartphone pulse oximeters.
Methods:
This was a prospective, observational study that involved noninvasive measurements of [Formula: see text] and heart rate with 3 devices: Masimo Radical-7, Kenek Edge with the Apple iPhone 6S, and the Samsung S8 smartphone. Ambulatory adult patients visiting our institution's pulmonary function lab for a 6-min walk test were eligible to participate in the study. Pretest and posttest results for each subject were obtained simultaneously using all 3 devices. All results were analyzed with the Spearman rho correlation test, and Bland-Altman plots were used to assess the agreement of measures between the devices.
Results:
Forty-seven subjects were enrolled in the study, with pulmonary hypertension (30%) and COPD (23%) being the 2 major diagnoses. The mean ± SD difference between the Masimo and Apple devices for pretest [Formula: see text] was 2.3 ± 2.4%, and the difference for posttest [Formula: see text] was 2.1 ± 3.9%. The mean difference between the Masimo and Samsung devices for pretest [Formula: see text] was 3.2 ± 2.8%, and the difference for posttest [Formula: see text] was 2.4 ± 3.5%. The number of subjects who were unable to obtain [Formula: see text] was higher with the Samsung device than with the Apple device in both pretest (14 of 47 vs 3 of 47) and posttest (17 of 47 vs 5 of 47). In contrast, the Masimo device was able to measure [Formula: see text] in all subjects.
Conclusions:
Smartphone pulse oximeters were unreliable compared to a hospital pulse oximeter. Further research is needed with evolving technology to better understand smartphone pulse oximetry. (ClinicalTrials.gov registration NCT03534271.).
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...