Compatibility of endoclips in the gastrointestinal tract with magnetic resonance imaging
Dong Yeol Shin1, Sumi Park2, Ain Kim3
1Division of Gastroenterology, Department of Internal Medicine, National Health Insurance Service Ilsan Hospital, 100 Ilsan-ro Ilsan-donggu, Goyang-si, Goyang, 10444, Korea.
Abstract:
There are no clear guidelines on the compatibility between endoclips that remain in the gastrointestinal (GI) tract and magnetic resonance imaging (MRI). The purpose of this study was to investigate the effect of 3T (T) MRI on endoclips placed in excised pig tissues. Two types of endoclips were assessed: Olympus EZ (HX-610-135L) and QuickClip Pro (HZ-202LR). We assessed tissue damage or perforation and detachment of endoclips under 3T MRI magnetic field. We also evaluated the magnitude of force required to detach the endoclips from the porcine tissue. We measured the magnetic force acting on the Olympus EZ clips. QuickClip Pro clips were used as a control in this study. There was no tissue damage and no detachment of the endoclips (Olympus EZ and QuickClip Pro) during 3T MRI. The force required to detach the Olympus EZ clips ranged from 0.9 to 3.0 N. The translational magnetic force acting on the endoclips was 3.18 × 10-3 N. Ex vivo experiments showed that the magnetic field generated by 3 MRI did not cause tissue damage or perforation and did not detach the endoclips. Olympus EZ clips and QuickClip Pro clips in the GI tract appear to be safe during 3T MRI.
Insights
This study found that Olympus EZ and QuickClip Pro endoclips are safe for use in the gastrointestinal tract during 3 Tesla Magnetic Resonance Imaging (MRI). No tissue damage or clip detachment occurred in ex vivo tests.
Area of Science:
- Medical Devices
- Gastroenterology
- Radiology
Background:
- Endoclips are used for gastrointestinal procedures.
- Compatibility of retained endoclips with Magnetic Resonance Imaging (MRI) is not well-defined.
- Patient safety during MRI is paramount.
Purpose of the Study:
- To evaluate the safety of commonly used endoclips during 3 Tesla (T) MRI.
- To assess potential tissue damage, perforation, or endoclip detachment.
- To determine the magnetic forces acting on endoclips in porcine tissue.
Main Methods:
- Ex vivo testing of Olympus EZ and QuickClip Pro endoclips in excised porcine tissue.
- Exposure to a 3T MRI magnetic field.
- Assessment of tissue integrity and endoclip stability.
- Measurement of detachment forces and magnetic forces.
Main Results:
- No instances of tissue damage or perforation were observed.
- Neither Olympus EZ nor QuickClip Pro endoclips detached during 3T MRI.
- Detachment forces for Olympus EZ clips ranged from 0.9 to 3.0 N.
- The translational magnetic force on endoclips was minimal (3.18 × 10⁻³ N).
Conclusions:
- 3T MRI is safe for patients with retained Olympus EZ and QuickClip Pro endoclips in the gastrointestinal tract.
- These endoclips demonstrate no adverse effects under 3T MRI conditions.
- Clinical guidelines can be informed by these ex vivo safety findings.
More Related Videos
09:29Fabrication and Implantation of Miniature Dual-element Strain Gages for Measuring In Vivo Gastrointestinal Contractions in Rodents.
Published on: September 18, 2014
09:42Murine Endoscopy for In Vivo Multimodal Imaging of Carcinogenesis and Assessment of Intestinal Wound Healing and Inflammation
Published on: August 26, 2014
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Endoscopic Procedures III: Video Capsule Endoscopy
Imaging Studies IV: Magnetic Resonance Imaging