Variation of Contrast Values for Myocardial Perfusion Imaging in Single-photon Emission Computed Tomography/Computed
Hazem M Tantawy1, Yasser G Abdelhafez2, Nadia L Helal3
1Department of Nuclear Medicine Technology, Inaya Medical Colleges, Riyadh, Saudi Arabia.
Objectives:
Single-photon emission computed tomography/computed tomography (SPECT/CT) hybrid systems have the advantage of performing various scans using the same imaging setting. Absorption and scattering of the gamma rays by the patient's body significantly affect images obtained from scintigraphy, especially in myocardial perfusion imaging. An important parameter for image quality in SPECT is image contrast which is defined as the difference in density between regions of the image corresponding to different levels of radioactive uptake in the patient. The objective of the study was to evaluate the influence of applying different correction methods on image contrast of myocardial SPECT/CT images.
Material And Methods:
A total of 114 patients, 43 females and 71 males, patient's raw data were processed and analyzed using attenuation correction (AC), scatter correction (SC), both attenuation and scatter correction together (ACSC), and no correction (NC). The short axis (coronal) slices resulted from the raw data reconstruction were chosen to perform the processing for hot and cold spheres for contrast values measurement. Statistical analysis was made for the measured contrast values for AC, SC, ACSC, and NC to determine the best image contrast.
Results:
When applying SC alone, it yields better contrast value (0.834), compared to AC (0.677) and ACSC (0.739). Both ACSC and AC had better image contrast compared to NC (0.592).
Conclusion:
The intercomparison study between the correction conditions indicates that the counts in SPECT/ CT are highly affected by all correction methods. The image contrast has been significantly improved by using SC, AC, and ACSC when compared with the NC image. Furthermore, SC is superior in the image contrast than the other correction conditions in the reconstruction of SPECT/CT MPI.
More Related Videos
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
06:29Cardiac Magnetic Resonance for the Evaluation of Suspected Cardiac Thrombus: Conventional and Emerging Techniques
Published on: June 11, 2019
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies for Cardiovascular System IV: CMRI