Related Experiment Video
Updated: Dec 6, 2025

Intra-Operative Neural Monitoring of Thyroid Surgery in a Porcine Model
Published on: February 11, 2019
Full percutaneous intraoperative neuromonitoring technique in remote thyroid surgery: Porcine model feasibility study
Tzu-Yen Huang1,2, Yi-Chu Lin1, Hsin-Yi Tseng1
1Department of Otorhinolaryngology-Head and Neck Surgery, Kaohsiung Medical University Hospital, Kaohsiung Medical University, Kaohsiung, Taiwan.
Background:
In remote thyroid surgery, a reliable intraoperative neuromonitoring (IONM) procedure is an important tool for reducing injury to recurrent laryngeal nerve (RLN). This study proposes an alternative or adjunct technique for performing full percutaneous (PC) IONM and confirms its feasibility in animal experiments.
Methods:
This prospective porcine model study enrolled four piglets with eight nerve sides. Evoked electromyography (EMG) was stimulated from PC ball-tip probe, and recorded from EMG endotracheal tube (ETT) and from PC paired long-needle electrodes on the perichondrium of the lateral aspect of thyroid cartilage.
Results:
In all RLNs and vagus nerves, typical laryngeal EMG waveforms were successfully evoked by PC probe stimulation and recorded by both ETT and PC needle electrodes.
Conclusions:
This study confirms the feasibility of the full PC IONM techniques in porcine model. However, further clinical studies are needed to compare the practicality of different remote-access approaches for thyroid surgery.
More Related Videos
11:46Investigation of the Electrophysiological and Thermographic Safety Parameters of Surgical Energy Devices During Thyroid and Parathyroid Surgery in a Porcine Model
Published on: October 13, 2022
05:25Identification and Protection of the Recurrent Laryngeal Nerve during Transoral Robotic Thyroidectomy
Published on: October 24, 2025