Ultrasound Assessment of Endothelial-Dependent Flow-Mediated Vasodilation of the Brachial Artery in Clinical Research
Published on: October 22, 2014
Standardization of penile hemodynamic evaluation through color duplex-doppler ultrasound
Felipe Carneiro1, Osmar Cassio Saito1, Eduardo P Miranda2
1Departamento de Radiologia, Universidade de São Paulo, São Paulo, SP, Brasil.
Insights
Color Duplex-Doppler Ultrasound (CDDU) of the penis aids erectile dysfunction treatment. This review promotes standardization for accurate CDDU conduction and interpretation, reducing variability in results.
Area of Science:
- Urology
- Vascular Imaging
- Sexual Medicine
Background:
- Color Duplex-Doppler Ultrasound (CDDU) of the penis is crucial for evaluating erectile function and guiding treatment decisions.
- Lack of a standardized CDDU procedure leads to significant variability in result interpretation.
- Standardization is needed to improve the reliability of penile vascular assessments.
Purpose of the Study:
- To promote standardization in the conduction of penile CDDU.
- To discuss fundamental principles for accurate penile CDDU.
- To reduce result-interpretation variability in erectile dysfunction assessments.
Main Methods:
- Penile CDDU begins with the flaccid penis, assessing the entire organ using B-mode imaging and a high-frequency linear transducer (7.5-18 MHz).
- Pharmacological erection is induced via intracavernous injection of vasodilating agents (e.g., prostaglandin E1).
- Serial measurements and penile rigidity assessments are performed throughout the session, considering factors like stimulation and re-dosing to optimize erection quality.
Main Results:
- Monitoring erection response using a hardness scale is essential.
- Optimizing scanning conditions during peak erection quality is recommended.
- Techniques to reduce psychogenic influences and achieve maximal erection aid accurate hemodynamic parameter assessment.
Conclusions:
- CDDU is a valuable diagnostic tool for erectile dysfunction.
- This guideline aims to standardize CDDU conduction and interpretation.
- Establishing uniformity in penile CDDU evaluations will enhance clinical decision-making.
Introduction:
The vascular evaluation of the erectile function through Color Duplex-Doppler Ultrasound (CDDU) of the penis can benefit the therapeutic decision-making process. Unfortunately, there is no standard procedure for CDDU conduction, a fact that results in high result-interpretation variability.
Objective:
The aims of this review are to promote greater standardization during CDDU of the penis and discuss the fundamental principles for its accurate conduction.
Methods:
CDDU is initially conducted with the penis in the flaccid state; the whole penis must be assessed (images at B mode) with a high-frequency linear transducer (7.5-18 MHz). Intracavernous injection of vasodilating agents (prostaglandin E1, papaverine, phentolamine) is performed to induce a rigid erection. Serial measurements at different times should be taken during the CDDU session and penile rigidity must be assessed in each evaluation.
Results:
It is important to monitor the erection response after the vasoactive agent (hardness scale), and scanning during the best-quality erection should be contemplated. Manual self-stimulation, audiovisual sexual stimulation (AVSS), and vasoactive agent re-dosing protocols must be taken into account to reduce the influence of psychogenic factors and to help the patient to get the hardest erection possible. Such measurements contribute to the maximal relaxation of the erectile tissue, so the hemodynamic parameters are not underestimated.
Conclusions:
CDDU is a relevant specialized tool to assess patients with erectile dysfunction; therefore, this guideline will help to standardize and establish uniformity in its conduction and interpretation, taking into consideration the complexity and heterogeneity of CDDU evaluations of the penis.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Urodynamic Studies: Uroflowmetry
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...