Synthetic MRI is not yet ready for morphologic and functional assessment of patellar cartilage at 1.5Tesla
C Vogrig1, J-S Louis2, F Avila1
1Guilloz Imaging Department, Central Hospital, Regional University Hospital Center of Nancy (CHRU-Nancy), 54000 Nancy, France.
Purpose:
The purpose of this study was to compare morphologic assessment and relaxometry of patellar hyaline cartilage between conventional sequences (fast spin-echo [FSE] T2-weighted fat-saturated and T2-mapping) and synthetic T2 short-TI inversion recovery (STIR) and T2 maps at 1.5T magnetic resonance imaging (MRI).
Method:
The MRI examinations of the knee obtained at 1.5T in 49 consecutive patients were retrospectively studied. There were 21 men and 28 women with a mean age of 45±17.7 (SD) years (range: 18-88 years). Conventional and synthetic acquisitions were performed, including T2-weighted fat-saturated and T2-mapping sequences. Two radiologists independently compared patellar cartilage T2-relaxation time on conventional T2-mapping and synthetic T2-mapping images. A third radiologist evaluated the patellar cartilage morphology on conventional and synthetic T2-weighted images. The presence of artifacts was also assessed. Interobserver agreement for quantitative variables was assessed using intraclass correlation coefficient (ICC).
Results:
In vitro, conventional and synthetic T2 maps yielded similar mean T2 values 58.5±2.3 (SD) ms and 58.8±2.6 (SD) ms, respectively (P=0.414) and 6% lower than the expected experimental values (P=0.038). Synthetic images allowed for a 15% reduction in examination time compared to conventional images. On conventional sequences, patellar chondropathy was identified in 35 patients (35/49; 71%) with a mean chondropathy grade of 4.8±4.8 (SD). On synthetic images, 28 patients (28/49; 57%) were diagnosed with patellar chondropathy, with a significant 14% difference (P=0.009) and lower chondropathy scores (3.7±4.9 [SD]) compared to conventional images. Motion artifacts were more frequently observed on synthetic images (18%) than on conventional ones (6%). The interobserver agreement was excellent for both conventional and synthetic T2 maps (ICC>0.83). Mean cartilage T2 values were significantly greater on synthetic images (36.2±3.8 [SD] ms; range: 29-46ms) relative to conventional T2 maps (31.8±4.1 [SD] ms; range: 26-49ms) (P<0.0001).
Conclusion:
Despite a decrease in examination duration, synthetic images convey lower diagnostic performance for chondropathy, greater prevalence of motion artifacts, and an overestimation of T2 values compared to conventional MRI sequences.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...