Low-dose CT with metal artifact reduction in arthroplasty imaging: a cadaveric and clinical study
Naveen Subhas1, Bong J Jun2, Parthiv N Mehta3
1Department of Diagnostic Radiology, Cleveland Clinic, 9500 Euclid Ave, Cleveland, OH, 44195, USA. subhasn@ccf.org.
Objective:
To determine whether a simulated low-dose metal artifact reduction (MAR) CT technique is comparable with a clinical dose MAR technique for shoulder arthroplasty evaluation.
Materials And Methods:
Two shoulder arthroplasties in cadavers and 25 shoulder arthroplasties in patients were scanned using a clinical dose (140 kVp, 300 qrmAs); cadavers were also scanned at half dose (140 kVp, 150 qrmAs). Images were reconstructed using a MAR CT algorithm at full dose and a noise-insertion algorithm simulating 50% dose reduction. For the actual and simulated half-dose cadaver scans, differences in SD for regions of interest were assessed, and streak artifact near the arthroplasty was graded by 3 blinded readers. Simulated half-dose scans were compared with full-dose scans in patients by measuring differences in implant position and by comparing readers' grades of periprosthetic osteolysis and muscle atrophy.
Results:
The mean difference in SD between actual and simulated half-dose methods was 2.42 HU (95% CI [1.4, 3.4]). No differences in streak artifact grades were seen in 13/18 (72.2%) comparisons in cadavers. In patients, differences in implant position measurements were within 1° or 1 mm in 149/150 (99.3%) measurements. The inter-reader agreement rates were nearly identical when readers were using full-dose (77.3% [232/300] for osteolysis and 76.9% [173/225] for muscle atrophy) and simulated half-dose (76.7% [920/1200] for osteolysis and 74.0% [666/900] for muscle atrophy) scans.
Conclusion:
A simulated half-dose MAR CT technique is comparable both quantitatively and qualitatively with a standard-dose technique for shoulder arthroplasty evaluation, demonstrating that this technique could be used to reduce dose in arthroplasty imaging.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography