Fracture Apparatus Design and Protocol Optimization for Closed-stabilized Fractures in Rodents
Published on: August 14, 2018
Variability of orthopedic physician fracture location identification: Implications for bone stimulator treatment
Scott Huff1, Joseph Henningsen1, Andrew Schneider1
1Department of Orthopedic Surgery, Miami Valley Hospital, Wright State University, 30, East Apple Street, Suite 2200, 45409 Dayton, OH, United States.
Introduction:
Aseptic non-union is a significant complication in approximately 5% of long-bone fractures. External bone stimulation treatment is often attempted before more invasive surgical interventions. Bone stimulators can have favorable results, but have a limited 1.7cm therapeutic radius. This study evaluated the accuracy by which clinicians locate a fracture on a cadaveric model. This has implications for the clinician's ability to accurately counsel patients on daily bone stimulator placement. Additionally, physicians (orthopedic attending surgeons and residents) were compared with pre-clinical (M1 and M2) medical students to evaluate if higher levels of training improved accuracy.
Hypothesis:
Orthopedic physicians and pre-clinical medical students will localize a radiographic fracture within 1.7cm less than 100% of the time, which represents the ideal consistency for patient care. Furthermore, orthopedic physicians will achieve a higher percentage accuracy than pre-clinical medical students.
Materials And Methods:
The sample included 20 orthopedic physicians and 16 pre-clinical medical students. Upper (radius) and lower (tibia) extremity cadaver models were prepared by inducing a single, transverse diaphyseal fracture. Plain reference radiographs of each model were obtained. Participants placed a radiopaque marker onto each model at the perceived fracture location, and radiographs were taken to document placement. Perpendicular marker-to-fracture distance was measured to the nearest mm along each bone's long axis using the PACS system.
Results:
Placement within the therapeutic radius was achieved by 70-80% of physicians, and 69-75% of medical students. In the remaining participants, improper placement distances were lower among physicians than among medical students (radius: 2.1±0.5 vs. 3.6±0.9cm, p=0.02; tibia: 2.6±0.5 vs 3.5±0.5cm, p=0.89).
Discussion:
In two cadaveric fracture models, up to 30% of orthopedic surgeons perceived a fracture location to be outside a bone stimulator's 1.7cm therapeutic radius. This finding suggests that physicians and their patients may benefit from additional methods for specifying the location of a non-union before commencing daily bone stimulator treatment.
Level Of Evidence:
Level IV, prospective cohort study-evidence from a well-designed prospective cohort study.