Related Experiment Video
Updated: Dec 6, 2025

Testing of all Six Semicircular Canals with Video Head Impulse Test Systems
Published on: April 18, 2019
Are gain values significantly altered by manual data selection when performing the video Head Impulse Test (v-HIT) on
Emil Riis Abrahamsen1,2, Regitze Kuhr Skals3, Dan Dupont Hougaard1,2
1Department of Otolaryngology, Head & Neck Surgery and Audiology, Aalborg University Hospital, Denmark.
Background:
It has not yet been tested whether averaged gain values and the presence of pathological saccades are significantly altered by manual data selection or if data selection only done by the incorporated software detection algorithms provides a reliable data set following v-HIT testing.
Objective:
The primary endpoint was to evaluate whether the averaged gain values of all six SCCs are significantly altered by manual data selection with two different v-HIT systems.
Method:
120 subjects with previously neither vestibular nor neurological disorders underwent four separate tests of all six SCCs with either EyeSeeCam® or ICS Impulse®. All v-HIT test reports underwent manual data selection by an experienced ENT Specialist with deletion of any noise and/or artifacts. Generalized estimating equations were used to compare averaged gain values based on unsorted data with averaged gain values based on the sorted data.
Results:
EyeSeeCam®: Horizontal SCCs: The estimate and the p-value (shown in parenthesis) for the right lateral SCC and the left lateral SCC were 0.00004 (0.95) and 0.00087 (0.70) respectively. Vertical SCCs: The estimate varied from -0.00858 to 0.00634 with p-values ranging from 0.31 to 0.78. ICS Impulse®: Horizontal SCCs: The estimate and the p-value for the right lateral SCC and the left lateral SCC were 0.00159 (0.18) and 0.00071 (0.38) respectively. Vertical SCCs: The estimate varied from 0.00217 to 0.01357 with p-values ranging from 0.00 to 0.17. Based upon the averaged gain value from the individual SCC being tested, 148 tests before and 127 after manual data selection were considered pathological.
Conclusion:
None of the two v-HIT systems revealed any clinically important effects of manual data selection. However, 21 fewer tests were considered pathological after manual data selection.
More Related Videos
07:24Using Eye-tracking to Assess the Relative Importance of Visual and Vestibular Input to Subcortical Motion Processing in the Roll Plane
Published on: August 22, 2025
10:12Three Dimensional Vestibular Ocular Reflex Testing Using a Six Degrees of Freedom Motion Platform
Published on: May 23, 2013
Related Concept Videos
The Vestibular System
Equilibrium and Balance