Determining Viral Disinfection Efficacy of Hot Water Laundering
Published on: June 21, 2022
Scalable, effective, and rapid decontamination of SARS-CoV-2 contaminated N95 respirators using germicidal
Importance:
Particulate respirators such as N95 masks are an essential component of personal protective equipment (PPE) for front-line workers. This study describes a rapid and effective UVC irradiation system that would facilitate the safe re-use of N95 respirators and provides supporting information for deploying UVC for decontamination of SARS-CoV-2 during the COVID19 pandemic.
Objective:
To assess the inactivation potential of the proposed UVC germicidal device as a function of time by using 3M 8211 - N95 particulate respirators inoculated with SARS-CoV-2.
Design:
A germicidal UVC device to deliver tailored UVC dose was developed and snippets (2.5cm2) of the 3M-N95 respirator were inoculated with 106 plaque-forming units (PFU) of SARS-CoV-2 and were UV irradiated. Different exposure times were tested (0-164 seconds) by fixing the distance between the lamp (10 cm) and the mask while providing an exposure of at least 5.43 mWcm-2.
Setting:
The current work is broadly applicable for healthcare-settings, particularly during a pandemic such as COVID-19.
Participants:
Not applicable. Main Outcome(s) and Measure(s): Primary measure of outcome was titration of infectious virus recovered from virus-inoculated respirator pieces after UVC exposure. Other measures included the method validation of the irradiation protocol, using lentiviruses (biosafety level-2 agent) and establishment of the germicidal UVC exposure protocol.
Results:
An average of 4.38x103 PFUml-1(SD 772.68) was recovered from untreated masks while 4.44x102 PFUml-1(SD 203.67), 4.00x102 PFUml-1(SD 115.47), 1.56x102 PFUml-1(SD 76.98) and 4.44x101 PFUml-1(SD 76.98) was recovered in exposures 2s,6s,18s and 54 seconds per side respectively. The germicidal device output and positioning was monitored and a minimum output of 5.43 mWcm-2 was maintained. Infectious SARS-CoV-2 was not detected by plaque assays (minimal level of detection is 67 PFUml-1) on N95 respirator snippets when irradiated for 120s per side or longer suggesting 3.5 log reduction in 240 seconds of irradiation.
Conclusions And Relevance:
A scalable germicidal UVC device to deliver tailored UVC dose for rapid decontamination of SARS-CoV-2 was developed. UVC germicidal irradiation of N95 snippets inoculated with SARS-CoV-2 for 120s per side resulted in 100% (3.5 log in total) reduction of virus. These data support the reuse of N95 particle-filtrate apparatus upon irradiation with UVC and supports use of UVC-based decontamination of SARS-CoV-2 virus during the COVID19 pandemic.
Related Concept Videos
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Hand hygiene
Hand washing...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Chemical Agents for Microbial Control