Related Experiment Video
Updated: Dec 6, 2025

Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Reliability of shoulder muscle stiffness measurement using strain ultrasound elastography and an acoustic coupler
Daisuke Shimoyama1, Hitoshi Shitara2, Noritaka Hamano1
1Department of Orthopaedic Surgery, Gunma University Graduate School of Medicine, 3-39-22 Showa, Maebashi, Gunma, 371-8511, Japan.
Purpose:
Abnormal increases in muscle tone can be caused by various musculoskeletal disorders. The objective of this study was to evaluate intratester and intertester reliabilities in measuring the stiffness of the shoulder muscles using strain ultrasound elastography (USE) and an acoustic coupler.
Method:
Tissue stiffness was measured in the trapezius muscle and in the supraspinatus muscle of healthy young volunteers.
Results:
The mean strain ratios measured by two experienced shoulder surgeons were significantly higher in the trapezius muscle than in the supraspinatus muscle (P < 0.001). Intratester reliability was rated as moderate to substantial for the trapezius muscle and substantial for the supraspinatus muscle. Intertester reliability was substantial for both muscles, with an intraclass correlation coefficient (2,1) of 0.62 [95% confidence interval (CI) 0.28-0.82] for the trapezius muscle and 0.69 (95% CI 0.40-0.86) for the supraspinatus muscle.
Conclusions:
We found substantial intratester and intertester reliabilities for the trapezius and supraspinatus muscles, suggesting that USE represents a promising modality for measuring the stiffness of shoulder muscles. However, the clinical application of this method will require the development of a device that can standardize the scanning technique to further increase the reliability.
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Measurements of Strain
Strain and Elastic Modulus