A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
Study of Three Interface Pressure Measurement Systems Used in the Treatment of Venous Disease
Gayani K Nandasiri1, Arash M Shahidi1, Tilak Dias1
1Advanced Textiles Research Group, School of Art and Design, Nottingham Trent University, Nottingham NG1 4GG, UK.
Abstract:
The aim of the publication is to report the accuracy, repeatability and the linearity of three commercially available interface pressure measurement systems employed in the treatment of venous disease. The advances in the treatment and management of chronic venous disease by compression therapy have led to considerable research interest in interface pressure measurement systems capable of measuring low-pressure ranges (10-60 mmHg). The application of a graduated pressure profile is key for the treatment of chronic venous disease which is achieved by using compression bandages or stockings; the required pressure profiles are defined in standards (BSI, RAL-GZ, or AFNOR) for different conditions. However, achieving the recommended pressure levels and its accuracy is still deemed to be a challenge. Thus, it is vital to choose a suitable pressure measurement system with high accuracy of interface pressure. The authors investigated the sensing performance of three commercially available different pressure sensors: two pneumatic based (AMI and PicoPress®) and one piezoresistive (FlexiForce®) pressure sensors, with extensive experimental work on their performance in terms of linearity, repeatability, and accuracy. Both pneumatic based pressure measurement systems have shown higher accuracy in comparison to the flexible piezoresistive pressure sensors.
Related Concept Videos
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Venous Thrombosis III: Interprofessional Care
Varicose Veins II: Diagnostic Studies and Interprofessional Care
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...