Early aortic dissection after aortic valve replacement with a Perceval sutureless bioprosthesis
Giosuè Falcetta1, Stefano Pratali1, Giacomo Ravenni1
1Section of Cardiac Surgery, Cardiothoracic and Vascular Department, University Hospital of Pisa, Via Paradisa, 2, 56124 Pisa, Italy.
Indian Journal of Thoracic and Cardiovascular Surgery
|October 16, 2020
Summary
Aortic dissection after aortic valve replacement using a Perceval bioprosthesis was observed. The dissection was not directly linked to the device, but similar prostheses may pose a risk.
Area of Science:
- Cardiovascular Surgery
- Medical Device Complications
- Aortic Diseases
Background:
- Aortic valve replacement is a common procedure.
- Sutureless bioprostheses offer potential advantages in surgical time and technique.
- Complications associated with novel device implantation require careful monitoring.
Observation:
- A patient developed ascending aorta dissection post-aortic valve replacement with a Perceval sutureless pericardial bioprosthesis.
- The intimal tear was identified at the aortotomy suture line during reoperation.
- The nitinol stent tips were distant from the dissected aortic segment.
Findings:
- The observed aortic dissection was not directly attributable to the Perceval bioprosthesis itself.
- The complication's location suggests it was related to the aortotomy closure rather than device impingement.
- A prior report links transcatheter prosthesis struts to aortic wall erosion.
Implications:
- While not directly caused by the device, this case highlights a potential, albeit rare, complication associated with sutureless aortic valve prostheses.
- Further investigation into the biomechanical interaction between sutureless devices and the aortic wall is warranted.
- This finding may inform future device design and surgical techniques to mitigate dissection risk.