Related Experiment Video
Updated: Dec 5, 2025

Multi-step Preparation Technique to Recover Multiple Metabolite Compound Classes for In-depth and Informative Metabolomic Analysis
Published on: July 11, 2014
Dixon's Q-test and Student's t-test to assess analog internal standard response in nonregulated LC-MS/MS bioanalysis
Deborah Barkley1, Panos Hatsis1, Jim Glick1
1NIBR, PK Sciences Bioanalytics, East Hanover, NJ 07936, USA.
Abstract:
Aim: In bioanalytical assays, analyte response is normalized to an internal standard response. When the internal standard works well, it compensates for processing and detection variability. However, in case the internal standard introduces additional variability, due to addition errors or other issues, scientists need to identify this. Results: A new method, using a Q-test for outliers and a t-test to compare internal standard response from different sample types, is applied to 15 cases. The results show that the Q-test/t-test, which uses confidence level rather than arbitrary cut-points, is more discerning of deviations compared with widely used methods. Conclusion: This work may improve the quality of and rationale for the internal standard response monitoring method.
Related Concept Videos
Comparing Experimental Results: Student's t-Test
Detection of Gross Error: The Q Test
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Therapeutic Drug Monitoring: Drug Analysis Methods