Related Experiment Video
Updated: Dec 5, 2025

Isolation of Quartz Grains for Optically Stimulated Luminescence OSL Dating of Quaternary Sediments for Paleoenvironmental Research
Published on: August 2, 2021
Temporal modelling of beryllium oxide ceramics' real-time OSL for dosimetry with a superficial 140 kVp X-ray beam
Levi Madden1, Alexandre Santos2, Enbang Li1
1Centre for Medical Radiation Physics, University of Wollongong, NSW 2522, Australia.
Abstract:
A new analysis method for the rtOSL of BeO ceramics is presented, using temporal curve fitting of an expected rtOSL signal to measured rtOSL signals. The presented technique does not require heavy signal averaging to determine the OSL bleaching correction associated with the ΔrtOSL method, reducing uncertainties in the post-correction rtOSL. The corrected rtOSL signal was demonstrated to be linear with dose, and dose-rate independent. The presented technique is expected to be applicable for many other dosimeters capable of the rtOSL technique. The presented technique achieved relative uncertainties in the corrected rtOSL between 3.4% and 6.5%. The initial measurements are promising, but uncertainties are required to be further improved upon before the technique can be used clinically.