Related Experiment Videos
Fluvo-metricell, a combined cell volume and cell fluorescence analyzer
Summary
A novel flow cytometer simultaneously measures cell volume and fluorescence using a unique transducer. This advancement allows for precise, coordinated analysis of cellular properties in a single measurement.
Area of Science:
- Biotechnology and Biomedical Engineering
- Cell Biology and Analysis
Background:
- Traditional cell analysis methods often require separate measurements for cell volume and fluorescence.
- Integrating these measurements can improve efficiency and accuracy in cellular studies.
Purpose of the Study:
- To introduce a new flow-through instrument for simultaneous measurement of cell volume and fluorescence.
- To demonstrate the instrument's capability for accurate and coordinated cellular analysis.
Main Methods:
- Utilizes the resistance pulse technique for cell volume measurement.
- Employs axial optical detection of fluorescence pulses from hydrodynamically focused cells.
- Incorporates a novel transducer to eliminate coordination issues between volume and fluorescence signals.
Main Results:
- The instrument accurately measures cell volumes between 4 and 1400 μm³ using a 70-micron orifice.
- Achieves coefficients of variation below 2% for fluorescence measurements.
- Demonstrates successful application through various practical examples and function tests.
Conclusions:
- The developed instrument offers a robust solution for simultaneous, high-precision cell volume and fluorescence analysis.
- This technology enhances cellular analysis by providing coordinated, multi-parametric data.
- The instrument is suitable for diverse applications in cell biology and biotechnology.