A Protocol of Manual Tests to Measure Sensation and Pain in Humans
Published on: December 19, 2016
Instrumental validity and intra/inter-rater reliability of a novel low-cost digital pressure algometer
Daniel Jerez-Mayorga1, Carolina Fernanda Dos Anjos2, Maria de Cássia Macedo2
1Facultad de Ciencias de la Rehabilitación, Universidad Andrés Bello, Santiago, Santiago, Chile.
Background:
Pain assessment is a key measure that accompanies treatments in a wide range of clinical settings. A low-cost valid and reliable pressure algometer would allow objective assessment of pressure pain to assist a variety of health professionals. However, the pressure algometer is often expensive, which limits its daily use in both clinical and research settings.
Objectives:
This study aimed to assess the instrumental validity, and the intra- and inter-rater reliability of an inexpensive digital adapted pressure algometer.
Methods:
A single rater applied 60 random compressions on a force platform. The pressure pain thresholds of 20 volunteers were collected twice (3 days apart) by two raters. The main outcome measurements were as follows: the maximal peak force (in kPa) and the pressure pain threshold (adapted pressure algometer vs. force platform). Cronbach's α test was used to assess internal consistency. The standard error of measurement provided estimates of measurement error, and the measurement bias was estimated with the Bland-Altman method, with lower and upper limits of agreement.
Results:
No differences were observed when comparing the compression results (P = 0.51). The validity and internal intra-rater consistencies ranged from 0.84 to 0.99, and the standard error of measurement from 0.005 to 0.04 kPa. Very strong (r = 0.73-0.74) to near-perfect (r = 0.99) correlations were found, with a low risk of bias for all measurements. The results demonstrated the validity and intra-rater reliability of the digitally adapted pressure algometer. Inter-rater reliability results were moderate (r = 0.55-0.60; Cronbach's α = 0.71-0.75).
Conclusion:
The adapted pressure algometer provide valid and reliable measurements of pressure pain threshold. The results support more widespread use of the pressure pain threshold method among clinicians.
More Related Videos
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
11:26Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression
Published on: December 10, 2014
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: