Toward Rapid Screening of Liver Grafts at the Operating Room Using Mid-infrared Spectroscopy
David Pérez-Guaita1, Marta Moreno-Torres2, Ramiro Jover2,3,4
1FOCAS Research Institute, Technological University Dublin, 13 Camden Row, D08CKP1 Dublin, Ireland.
Analytical Chemistry
|October 21, 2020
Summary
Fourier transform infrared (FTIR) spectroscopy offers a rapid, in-situ method for assessing liver steatosis in transplant grafts. This technique accurately quantifies fat content, aiding surgeons in graft selection and improving patient outcomes.
Area of Science:
- Biomedical Engineering
- Spectroscopy
- Transplantation Science
Background:
- Liver graft steatosis estimation is critical for transplantation success, as high fat levels increase graft failure risk.
- Current assessment methods (visual inspection, palpation, histology) are subjective and time-consuming.
- A need exists for rapid, in-situ diagnostic tools for pre-transplant graft evaluation.
Purpose of the Study:
- To evaluate an attenuated total reflection-Fourier transform infrared (ATR-FTIR) spectroscopic method for in-situ liver steatosis assessment.
- To determine if ATR-FTIR can provide fast, quantitative lipid content analysis in human liver biopsies.
- To explore ATR-FTIR as an alternative to traditional pathological examination for liver grafts.
Main Methods:
- Analysis of 20 human liver biopsies using ATR-FTIR spectroscopy.
- Comparison of ATR-FTIR data with histological analysis, biochemical lipid analysis, and lipidomic profiling (UPLC-ESI(+/-)TOFMS).
- Assessment of ATR-FTIR method's compatibility with operating room requirements.
Main Results:
- Strong correlations were observed between ATR-FTIR data and clinical/lipidomic information.
- ATR-FTIR effectively captured global liver lipid composition and quantified triacylglycerol content.
- The method demonstrated potential for in-situ grading of liver steatosis.
Conclusions:
- ATR-FTIR spectroscopy shows promise as a fast, quantitative tool for in-situ liver steatosis assessment in the operating room.
- This spectroscopic method could serve as a valuable alternative to subjective pathological examinations.
- Further validation is required, but results support ATR-FTIR's applicability in liver transplantation.