Assessing an alpha-defensin lateral flow device for diagnosing septic arthritis: reporting on a false-negative case

Atsushi Narita1, Akemi Suzuki1, Taku Nakajima1

  • 1Department of Orthopaedic Surgery, Faculty of Medicine, Yamagata University, Yamagata, Japan.

Insights

Alpha-defensin (αD) shows potential for diagnosing joint infections but is not definitive for septic arthritis. Further studies are needed to confirm its diagnostic accuracy for this condition.

Area of Science:

  • Biochemistry
  • Immunology
  • Clinical Diagnostics

Background:

  • Alpha-defensin (αD), an antimicrobial peptide, is released by neutrophils and has been proposed as a diagnostic biomarker in synovial fluid.
  • Previous studies indicate αD is a reliable biomarker for diagnosing periprosthetic joint infection (PJI).

Observation:

  • This study investigated the utility of αD for diagnosing septic arthritis, a condition similar to PJI.
  • The Synovasure® αD detection lateral flow device was used to assess suspected septic arthritis cases.
  • The study identified both a false-positive and a false-negative result.

Findings:

  • A false-negative case involved infection by *Staphylococcus capitis*, a coagulase-negative and potentially low-virulence organism.
  • A false-positive case was diagnosed as seronegative rheumatoid arthritis with calcium pyrophosphate depositions, suggesting αD may be elevated in conditions with neutrophil mobilization.
  • The αD lateral flow device alone proved insufficient for definitive septic arthritis diagnosis.

Implications:

  • While αD may serve as an additional indicator in difficult PJI diagnoses, its utility in septic arthritis remains inconclusive.
  • Larger prospective clinical studies are necessary to establish the definitive role and accuracy of αD in diagnosing septic arthritis.