Assessing an alpha-defensin lateral flow device for diagnosing septic arthritis: reporting on a false-negative case
Atsushi Narita1, Akemi Suzuki1, Taku Nakajima1
1Department of Orthopaedic Surgery, Faculty of Medicine, Yamagata University, Yamagata, Japan.
Abstract:
Alpha-defensin (αD), an antimicrobial peptide released by neutrophils in response to bacterial pathogens, was proposed as a novel diagnostic biomarker in synovial fluid. Several reports have shown that αD can serve as a reliable biomarker in the diagnosis of periprosthetic joint infection (PJI). We assessed whether αD could also serve to diagnosis of septic arthritis, a similarly difficult to diagnose PJI. To our knowledge, besides PJI, few reports exist assessing the utility of αD for septic arthritis. We have attempted to diagnose several cases of suspected septic arthritis using the Synovasure® αD detection lateral flow device. We report a false-positive case and a false-negative case. The false-negative case we experienced was caused by Staphylococcus capitis, which is coagulase-negative, and possibly represents a low virulence micro-organism infection. The false-positive case was ultimately diagnosed as seronegative rheumatoid arthritis and possessed calcium pyrophosphate depositions. False positives have been suggested to occur in conditions where neutrophils are mobilised. As for PJI, in cases where diagnosis is difficult, αD can be an additional diagnostic indicator. However, making a definitive diagnosis using the αD lateral flow device alone was found to be difficult. The utility of αD in assessing septic arthritis is inconclusive; therefore, larger prospective clinical studies should be considered for a better assessment.
Insights
Alpha-defensin (αD) shows potential for diagnosing joint infections but is not definitive for septic arthritis. Further studies are needed to confirm its diagnostic accuracy for this condition.
Area of Science:
- Biochemistry
- Immunology
- Clinical Diagnostics
Background:
- Alpha-defensin (αD), an antimicrobial peptide, is released by neutrophils and has been proposed as a diagnostic biomarker in synovial fluid.
- Previous studies indicate αD is a reliable biomarker for diagnosing periprosthetic joint infection (PJI).
Observation:
- This study investigated the utility of αD for diagnosing septic arthritis, a condition similar to PJI.
- The Synovasure® αD detection lateral flow device was used to assess suspected septic arthritis cases.
- The study identified both a false-positive and a false-negative result.
Findings:
- A false-negative case involved infection by *Staphylococcus capitis*, a coagulase-negative and potentially low-virulence organism.
- A false-positive case was diagnosed as seronegative rheumatoid arthritis with calcium pyrophosphate depositions, suggesting αD may be elevated in conditions with neutrophil mobilization.
- The αD lateral flow device alone proved insufficient for definitive septic arthritis diagnosis.
Implications:
- While αD may serve as an additional indicator in difficult PJI diagnoses, its utility in septic arthritis remains inconclusive.
- Larger prospective clinical studies are necessary to establish the definitive role and accuracy of αD in diagnosing septic arthritis.
More Related Videos
08:01Detection of Invasive Pulmonary Aspergillosis in Haematological Malignancy Patients by using Lateral-flow Technology
Published on: March 22, 2012
08:26Development and Validation of an Ultrasensitive Single Molecule Array Digital Enzyme-linked Immunosorbent Assay for Human Interferon-α
Published on: June 14, 2018
Related Concept Videos
Urine Studies II: Urine Culture and Sensitivity Test
Acute Pyelonephritis II: Diagnostic Studies and Management