A Silicon-tipped Fiber-optic Sensing Platform with High Resolution and Fast Response
Published on: January 7, 2019
A novel fiber-optic based 0.014″ pressure wire: Designs of the OptoWire™, development phases, and the O2 first-in-man
Paola Ulacia1,2, Goran Rimac1, Sébastien Lalancette3
1Department of Cardiology, Research Center, Quebec Heart-Lung Institute, Quebec, Quebec, Canada.
Objectives:
To review the technical limitations of available pressure-wires, present the design evolution of a nitinol fiber-optic pressure wire and to summarize the First-in-Man (FIM) O2 pilot study results.
Background:
Despite increasing use of physiology assessment of coronary lesions, several technical limitations persist. We present technical details, design evolution and early clinical results with a novel 0.014" nitinol fiber-optic based pressure-wire.
Methods And Results:
The 0.014' OptoWire™ (Opsens Medical, Quebec, Canada) was designed to combine improved handling properties compared to standard pressure-wires and to offer extremely reliable pressure recording and transmission due to fiber-optic properties compared to piezo-electric sensors and electrical wires. In vitro assessment showed that OptoWire™ steerability, pushability and torquability properties were closer to regular PCI wires than standard electrical pressure wires. In the First-in-Man O2 study, 60 patients were recruited at 2 centers in Canada. A total of 103 lesions were assessed with the OptoWire™ and OptoMonitor™, 75 lesions at baseline and 28 lesions post-PCI (without disconnection). In all crossed lesions (n = 100, 97%), mean Pd/Pa and FFR could be adequately measured. In 11 cases assessed successively with OptoWire™ and Aegis™ (Abbott Vascular, USA) bland-Altman analysis showed a mean difference of 0.002 ± 0.052 mmHg (p = .91) for Pd/Pa and 0.01 ± 0.06 for FFR calculation (p = .45). There was no device-related complication. Upon these initial results, several design changes aimed to improve overall performance including torquability, stiffness, resistance to kink and pressure drift were completed.
Conclusion:
The novel 0.014" fiber-optic OptoWire™ provides superior wire handling with reduced risk of pressure drift allowing reliable pre- and post-PCI physiology assessment.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Pressure Gauges