Related Experiment Video
Updated: Dec 4, 2025

Assessment of Bone Fracture Healing Using Micro-Computed Tomography
Published on: December 9, 2022
Improved diagnostic confidence in evaluating bone non-union using virtual monochromatic dual-energy CT
J C E Donders1, R H H Wellenberg2, G J Streekstra3
1Department of Orthopedic Surgery, Amsterdam UMC, Location AMC, Amsterdam, the Netherlands.
Purpose:
The aim of this study was to determine whether virtual monochromatic dual-energy CT imaging improves the evaluation of suspected non-union of the appendicular skeleton treated with titanium or stainless steel intramedullary nails and plates.
Method:
Forty-one patients with a clinical suspected non-union with hardware in place were included and scanned on a dual-source CT-scanner using 100/Sn150kVp. Images including titanium hardware were extracted at 130 keV. Images including stainless steel hardware were extracted at 150 keV. Monochromatic 70 keV images served as reference. Non-union confirmed during revision surgery was used as gold standard. A musculoskeletal radiologist and orthopedic trauma surgeon evaluated images on image quality, degree and location of consolidation, non-union type and diagnostic confidence.
Results:
Likert scores with respect to image quality improved from 0.88 to 1.83 (p < 0.001) in high (130 and 150) keV images. High keV images reduced the number of false negative non-unions based on consolidation grade with 5% (p = 0.283). Agreement between observers regarding location of consolidation and non-union type did not improve in 130 and 150 keV images. Diagnostic confidence improved from 1.43 to 2.37 in high keV images compared to 70 keV images (p < 0.001). Overall diagnostic confidence was higher in intramedullary nails than plates (p < 0.05).
Conclusions:
Use of virtual monochromatic 130 and 150 keV dual-energy CT compared to 70 keV images improves the evaluation of suspected non-union of the appendicular skeleton treated with titanium or stainless steel intramedullary nails and plates.
More Related Videos
18:40Combined In vivo Optical and µCT Imaging to Monitor Infection, Inflammation, and Bone Anatomy in an Orthopaedic Implant Infection in Mice
Published on: October 16, 2014
08:39Longitudinal Micro-Computed Tomography Image Analysis for User-Defined Region of Interest in Critical-Sized Bone Defects
Published on: June 24, 2025
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...