Abstract:
The HeartMate 3 [HM3 (Abbott, Abbott Park, Illinois)] is a left-ventricular assist device (LVAD) with excellent clinical results. Outflow graft occlusion as a complication secondary to outflow graft twisting was reported to occur in 1.6% within the MOMENTUM 3 trial. The anti-twist metal clip or modified bend relief is made to prevent this complication currently, however, there remain large numbers of early implanted HM3 which may develop this complication. There are limited reports illustrating diagnosis, surgical repair, and post-repair hemodynamic changes of these complications. Thus, we present a case of successful diagnosis and surgical repair of an outflow graft twist. Simple procedure through thoracotomy without cardiopulmonary bypass provides significant immediate hemodynamic improvement.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
10:05Testing the Efficacy of Pharmacological Agents in a Pericardial Target Delivery Model in the Swine
Published on: July 7, 2016
Related Concept Videos
Cardiomyopathy III: Hypertrophic Cardiomyopathy
Heart Failure V: Medical Management
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Heart Failure VI: Adjunct Therapies