Real-Time Dynamic Navigation System for the Precise Quad-Zygomatic Implant Placement in a Patient with a Severely Atrophic Maxilla
Published on: October 18, 2021
Comparative accuracy of cone-beam CT and conventional multislice computed tomography for real-time navigation in
Baoxin Tao1, Yihan Shen1, Yuanyuan Sun1
1Department of Second Dental Center, Shanghai Ninth People's Hospital, College of Stomatology, Shanghai Jiao Tong University School of Medicine; National Clinical Research Center for Oral Diseases, Shanghai Key Laboratory of Stomatology & Shanghai Research Institute of Stomatology, Shanghai, China.
Background:
Cone-beam computed tomography (CBCT) and conventional multislice CT (MSCT) are both used in zygomatic implant navigation surgery but the superiority of one technique versus the other remains unclear.
Purpose:
This study compared the accuracy of CBCT and MSCT in zygomatic implant navigation surgery by calculating the deviations of implants.
Material And Methods:
Patients with severely atrophic maxillae were classified into two groups according to the use of CBCT- or MSCT-guided navigation system. The entry and apical distance deviation, and the angle deviation of zygomatic implants were measured on fused operation images. A linear effect model was used for analysis, with statistical significance set at P < .05.
Results:
A total of 72 zygomatic implants were inserted as planned in 23 patients. The comparison of deviations in CBCT and MSCT groups showed a mean (± SD) entry deviation of 1.69 ± 0.59 mm vs 2.04 ± 0.78 mm (P = .146), apical deviation of 2 ± 0.68 mm vs 2.55 ± 0.85 (P < .001), and angle deviation of 2.32 ± 1.02° vs 3.23 ± 1.21° (P = .038).
Conclusion:
Real-time zygomatic implant navigation surgery with CBCT may result in higher values for accuracy than MSCT.