Related Experiment Video
Updated: Dec 3, 2025

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Tetracaine from urethral ointment causes false positive amphetamine results by immunoassay
Robin Wijngaard1, Marina Parra-Robert1, Lourdes Marés1
1Pharmacology and Toxicology Laboratory, Service of Biochemistry and Molecular Genetics, Biomedical Diagnostic Center (CDB), University of Barcelona, Hospital Clínic, Barcelona, Spain.
Background:
Amphetamine urine drug screening by immunoassay is prone to cross-react with other compounds leading to false positive results. Tetracaine is a local anesthetic drug used in the clinical setting as an ointment during urinary catheterization. In our laboratory, tetracaine is often detected by gas chromatography-mass spectrometry in the urine of patients admitted in the emergency department with false positive amphetamine results. The objectives of this study were to investigate if there was cross-reactivity to tetracaine in an amphetamine immunoassay and to retrospectively evaluate the potential contribution of tetracaine to false positive amphetamine results.
Methods:
An interference study was conducted using negative urine samples spiked with increasing concentrations of tetracaine hydrochloride and analyzed with the CEDIA Amphetamine/Ecstasy immunoassay. Retrospectively, urine samples of patients which yielded positive amphetamine immunoassay results and were analyzed by gas chromatography-mass spectrometry were reviewed (n = 417). The presence of tetracaine and/or other drugs by gas chromatography-mass spectrometry were gathered.
Results:
Tetracaine caused false positive amphetamine results by immunoassay (cut-off 1000 µg/L) with a concentration of above 40 mg/L. Retrospective analysis of all positive amphetamine immunoassay samples showed that in 45 out of the 417 (10.8%) urine samples no amphetamine-like derivative was identified by gas chromatography - mass spectrometry. In 37 out of 45 (82.2%) of these false positive cases tetracaine was detected, of whom 59.5% (22/37) had an estimated tetracaine concentration of ≥40 mg/L.
Conclusions:
This study confirmed the interference of tetracaine in the CEDIA Amphetamine/Ecstasy immunoassay and that tetracaine may have contributed to around 80% of the false positive amphetamine cases in the urine samples of patients admitted to the emergency department at our institution.
More Related Videos
06:06Color Spot Test As a Presumptive Tool for the Rapid Detection of Synthetic Cathinones
Published on: February 5, 2018
04:19Author Spotlight: An Improved Technique for Trimethylamine Detection in Animal-Derived Medicine by Headspace Gas Chromatography-Tandem Quadrupole Mass Spectrometry
Published on: March 10, 2023
Related Concept Videos
Adrenergic Agonists: Indirect-Acting Agents
One mechanism involves depleting stored catecholamines by displacing them from synaptic vesicles. These agents, known as "displacers," are transported into vesicles at the expense of noradrenaline. Examples include amphetamine and tyramine, which lack a catechol moiety, resulting in prolonged action, improved oral...
Drugs Affecting Neurotransmitter Release or Uptake