3D fast low-angle shot (FLASH) technique for 3T contrast-enhanced brain MRI in the inpatient and emergency setting:
Sohil H Patel1, Prem P Batchala2, Kellan Schallert2
1Department of Radiology and Medical Imaging, University of Virginia Health, PO Box 800170, Charlottesville, VA, 22908, USA. shp4k@virginia.edu.
Purpose:
To retrospectively evaluate the diagnostic performance of a 1-min contrast-enhanced 3D-FLASH pulse sequence for detecting intracranial enhancing lesions compared to standard contrast-enhanced 3D-MPRAGE pulse sequence.
Methods:
Contrast-enhanced 3D-FLASH (acquisition time 49 s) and contrast-enhanced 3D-MPRAGE (4 min 35 s) pulse sequences were performed consecutively in 110 inpatient/emergency department 3T MRI brain examinations and analyzed by two independent neuroradiologist readers. For each sequence, the readers recorded (1) number of enhancing intracranial lesions; (2) intracranial susceptibility artifact (presence or absence; mm depth of intracranial signal loss); and (3) motion artifact (none, mild, moderate, severe). Inter and intra-reader agreement and reader accuracy relative to a reference standard were determined, and sequence comparison with respect to susceptibility and motion artifacts was performed.
Results:
There was substantial intra-reader, inter-sequence agreement [reader 1, κ = 0.70 (95% CI: [0.60, 0.81]); reader 2, κ = 0.70 (95% CI: [0.59, 0.82])] and substantial intra-sequence, inter-reader agreement [3D-MPRAGE assessment, κ = 0.76 (95% CI: [0.66, 0.86]); 3D-FLASH assessment, κ = 0.86 (95% CI: [0.77, 0.94]) for detection of intracranial enhancing lesions. For both readers, the diagnostic accuracy of 3D-FLASH and 3D-MPRAGE was similar (3D-MPRAGE: 86.4 and 88.1%; 3D-FLASH: 88.2 and 84.5%), with no inter-sequence diagnostic accuracy discordancy between the sequences for either reader. 3D-FLASH was associated with less susceptibility artifact (p < 0.001 both readers) and less motion artifact (p < 0.001 both readers).
Conclusion:
On 3T brain MRI in the inpatient and emergency department setting, 1-min 3D-FLASH pulse sequence achieved comparable diagnostic performance to 4.5 min 3D-MPRAGE pulse sequence for detecting enhancing intracranial lesions, with reduced susceptibility and motion artifacts.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...