Universal response method for the quantitative analysis of multi-components in josamycin and midecamycin using liquid
Guijun Liu1, Bingqi Zhu2, Xiaojuan Ren1
1Zhejiang University of Technology, Hangzhou 310014, China.
Abstract:
Josamycin and midecamycin are consisted of three groups of components with different ultraviolet maximum absorption wavelengths (λmax), which are 231 nm, 280 nm and 205 nm. The quantitative analysis of all these components is challengeable due to the absence of the respective reference substances. To address this problem, universal and reliable methods were developed using high performance liquid chromatography coupled with charged aerosol detector (HPLC-CAD) for the quantitative analysis of components in josamycin and midecamycin. The chromatographic conditions and CAD parameters setting were optimized. Subsequently, the components were identified using HPLC coupled with ion trap/time-of-flight mass spectrometry (IT/TOF MS). The developed methods were validated by assessing linearity, limit of quantitation (LOQ), accuracy, precision and robustness. Good separations were achieved for all components and the adjustment of the filter valve and power function value efficiently improved sensitivity. The developed methods were more comprehensive than current HPLC-UV method. The experimental results demonstrated good linearity with coefficients of determination (R2) greater than 0.999 in the range of 0.002-0.30 mg mL-1. The limits of detection (LOD) were ranging from 1.8 to 2.0 μg·mL-1. The intra-day and inter-day RSD values were less than 2.0 % (n = 6) and 5.6 % (n = 9) respectively. The recoveries were 95.0 %-124.0 % at the spiked concentration levels of 0.05 %, 0.50 %, 0.10 % and 2.5 % with relative standard deviations (RSDs, n = 3) lower than 2.0 %. Finally, the developed methods were successfully applied to the quantitative analysis of minor components and used main components (leucomycin A3 and midecamycin A1) as alternative reference substance of minor components. The overall results demonstrated that the HPLC-CAD was a good alternative for the quantitative analysis of multi-components in 16-membered macrolides.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Analyzers: Common Types
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
High-Performance Liquid Chromatography: Types of Detectors