Indices for the evaluation of glandular dose heterogeneity in full-field digital mammography
Sae Shinohara1,2, Fujio Araki3, Megumi Maeda4
1Graduate School of Health Sciences, Kumamoto University, 4-24-1 Kuhonji, Chuo-ku, Kumamoto 862-0976, Japan.
Abstract:
This study aims to evaluate the indices of glandular dose heterogeneity in full-field digital mammography. The distributions of GD in a breast phantom with a skin layer of 4 mm were determined using the Monte Carlo method with simulated x-ray fluence spectra. First, the GD to air kerma (GD/Kair) volume histogram was obtained from the GD distributions, which were indicated by the glandular volume (%) as a function of GD/Kair. The GD indices, namely, the maximum glandular dose (GD2%) and glandular volume percentage (%) receiving at least the mean glandular dose (MGD) (VMGD) were calculated from the GD/Kairvolume histogram. Next, the scatter plots of GD2%/MGD andVMGDwere drawn as functions of the normalised mean glandular dose (DgN). Finally, (GD2%)and (VMGD)were obtained from the relationship between the GD indices and DgN for 596 clinical irradiation cases based on individual irradiation conditions. The values of GD2%/MGD were more affected by breast thickness than glandularity and tube voltage, and they decreased according to the power law of DgN for all the target/filter combinations. The values ofVMGDwere proportional to DgN and decreased with increase in the compressed breast thickness. The values of (MGD)and (GD2%)for 596 clinical irradiation cases were estimated to range from 0.6-3.0 mGy to 1.1-7.0 mGy, respectively, and (VMGD)was in the range 32%-48%. (GD2%)and (VMGD)are mainly affected by breast thickness. These indices are useful for the evaluation of glandular dose heterogeneity in mammography.