Related Experiment Video
Updated: Dec 3, 2025

Optical Coherence Tomography Based Biomechanical Fluid-Structure Interaction Analysis of Coronary Atherosclerosis Progression
Published on: January 15, 2022
Computational fluid dynamics simulations of flow distribution and graft designs in apicoaortic bypass
Takashi Sasaki1, Hitomi Ueda2, Keiichi Itatani3
1Department of Cardiovascular Surgery, Nippon Medical School, 1-1-5, Sendagi, Bunkyo-ku, Tokyo, 113-8603, Japan. t-sasaki0@nms.ac.jp.
Objective:
Apicoaortic bypass has double outlets and its graft design is similar to that of a left ventricular assist device (LVAD). The left ventricular apex to the descending aorta (LV-DsAo) bypass is widely used in apicoaortic bypass. In contrast, the left ventricular apex to the ascending aorta (LV-AsAo) bypass is standard in LVAD surgery. This study aimed to evaluate the graft designs of apicoaortic bypass and their effects on flow distribution and energy loss (EL).
Methods:
A simulation study using computational fluid dynamics was performed on the geometry and hemodynamics data obtained from a 30-year-old patient who underwent a LV-DsAo bypass. The ratio of the cardiac output (CO) through the ascending aorta (AsAo) and apicoaortic conduit was set at 50:50, 30:70, and 10:90. Regional blood flow (RBF) and EL were calculated for the different distribution ratios. As an alternative to the LV-DsAo bypass, a virtual LV-AsAo bypass surgery was performed, and each parameter was compared with that of the LV-DsAo bypass.
Results:
At a distribution ratio of 50:50, the RBF to the head and EL were 16.4% of the total CO and 62.0 mW in the LV-DsAo bypass, and 32.3% and 81.5 mW in the LV-AsAo bypass, respectively. The RBF to the head decreased with the CO through the AsAo in the LV-DsAo bypass, but it was constant in the LV-AsAo bypass. The EL increased inversely with the CO through the AsAo in both graft designs.
Conclusion:
The regional blood flow distribution was different, but the trend of the EL which increased inversely with the CO through the AsAo was similar between the LV-DsAo and LV-AsAo bypasses.
More Related Videos
12:00Evaluation of a Novel Laser-assisted Coronary Anastomotic Connector - the Trinity Clip - in a Porcine Off-pump Bypass Model
Published on: November 24, 2014
07:21Microsurgical Creation of Giant Bifurcation Aneurysms in Rabbits for the Evaluation of Endovascular Devices
Published on: September 8, 2023