Ultrasonography of the Adult Male Urinary Tract for Urinary Functional Testing
Published on: August 14, 2019
Accuracy of Measuring Bladder Volumes With Ultrasound and Bladder Scanning
Marilyn Schallom1, Donna Prentice1, Carrie Sona1
1Marilyn Schallom is director of research, Donna Prentice is a research scientist, Kara Vyers is a research coordinator, and Cassandra Arroyo is a statistician in the Department of Research and Carrie Sona is a clinical nurse specialist in the Department of Surgical Services, Barnes-Jewish Hospital, St Louis, Missouri.
Background:
Removal of urinary catheters depends on accurate noninvasive measurements of bladder volume. Patients with acute kidney injury often have low bladder volumes/ascites, possibly causing measurement inaccuracy.
Objective:
To evaluate the accuracy of bladder volumes measured with bladder scanning and 2-dimensional ultrasound (US) compared with urinary catheterization among different types of clinicians.
Methods:
Prospective correlational descriptive study of 73 adult critical care patients with low urine output receiving hemodialysis or unable to void. Bladder volumes were independently measured by (1) a physician and an advanced practice registered nurse using US, (2) an advanced practice registered nurse and a bedside nurse using bladder scanning, and (3) urinary catheterization (cath). Bland-Altman and χ2 analyses were conducted.
Results:
Mean (SD) cath volume was 171.7 (269.7) mL (range, 0-1100 mL). Abdominal fluid was observed in 28% of patients. Bias was -1.3 mL for US vs cath and 3.3 mL for bladder scanning vs cath. For patients with abdominal fluid and cath volume less than 150 mL, decisions to not catheterize patients were accurate more often when based on US measurements (97%-100%) than when based on bladder scanning measurements (86%-89%; P = .02). In patients with cath volume of 300 mL or more, decisions to catheterize patients were accurate more often when based on bladder scanning measurements (94%-100%) than when based on horizontal US measurements (50%-56%; P = .001).
Conclusions:
Bladder volume can be measured accurately with bladder scanning or US, but abdominal fluid remains a confounding factor limiting accuracy of bladder scanning.
Related Concept Videos
Imaging Studies II: Ultrasonography
Urodynamic Studies: Uroflowmetry
Imaging Studies VI: Voiding Cystourethrography and Cystography
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Urinary Bladder
In males, the bladder is situated in front of the rectum, while in females, it is positioned anterior to the vagina and uterus. The bladder floor contains an inverted triangular area called the trigone, defined by the two ureteric...
Imaging Studies IV: Magnetic Resonance Imaging