Evaluations for Determination of Optimum Shields in Nuclear Medicine
Parvaresh R1, Jalili M2, Haghparast A3
1MSc, Medical Physics Department, Faculty of Medicine, Kermanshah University of Medical Sciences, Kermanshah, Iran.
Background:
131I source is widely used in the treatment of hyperthyroidism and thyroid cancers. 131I emits both beta and gamma-rays. Radiation protection is considered for gamma rays emitted by 131I. It seems no special shield against 131I source to be designed.
Objective:
This research aims to evaluate determination of optimum shields in nuclear medicine against 99Tcm and 131I sources by dosimetric method. Additionally, Monte Carlo simulation was used to find the optimum thickness of lead for protection against 131I source.
Material And Methods:
This is an experimental research in the field of radiation protection. A calibrated model of GraetzX5C Plus dosimeter was used to measure exposure rates passing through the shields. The efficiency of the shields was evaluated against 99Tcm and 131I. Furthermore, Monte Carlo simulation was used to find the optimum thickness of lead for protection against 131I source.
Results:
The findings of the dosimetric method show that the minimum and maximum efficiencies obtained by the lead apron with lead equivalent thickness of 0.25 mm and the syringe holder shields with thickness of 0.5 mm lead were 50.86% and 99.50%, respectively. The results of the simulations show that the minimum and maximum efficiencies obtained by lead thicknesses of 1 mm and 43 mm were 19.36% and 99.79%, respectively.
Conclusion:
The optimum shields against 99Tcm are the syringe holder shield, the tungsten syringe shield, and the lead partition, respectively. Furthermore, based on simulations, the thicknesses of 11-28 mm of lead with efficiencies between 90.6% to 99% are suggested as the optimum thicknesses to protect against 131I source.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
09:03Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
Related Concept Videos
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Biological Effects of Radiation
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET