Related Experiment Video
Updated: Dec 2, 2025

Rapid Point-of-Care Assay of Enoxaparin Anticoagulant Efficacy in Whole Blood
Published on: October 12, 2012
Effect of anticoagulant adjustment on prothrombin time test using two different PT reagents in patients with elevated
Vanessa M Silva1, Denise C Rezende1, Edna S Garcia1
1Clinical Laboratory Instituto do Coracao, Hospital das Clinicas HCFMUSP, Faculdade de Medicina, Universidade de Sao Paulo, Sao Paulo, SP, Brazil.
Abstract:
The recommendations for adjustment of citrate volume in sample tubes with high hematocrit (Ht) are based on indirect studies of underfilled tubes or artificially constructed Ht values. The aim of this study was to evaluate the effect of citrate volume adjustment in sample tubes from patients with hematocrit >55% using two different prothrombin time (PT) tests.
Methods:
Paired citrate-adjusted and unadjusted blood specimens were obtained from 181 patients from the pulmonary hypertension ambulatory with high Ht values and on warfarin therapy. The samples were tested using recombinant human tissue factor (RTF) and reagents extracted from rabbit brain (HS Plus). The results are expressed as the international normalized ratio (INR). The correlation and percent change (% change) between sample pairs were calculated.
Results:
INR-RTF results from adjusted and unadjusted citrate blood specimens showed a strong correlation (R2 = 0.8226, p < 0.0001). The INR median was 2.25 (95% CI 2.10 to 2.41) for citrate-adjusted samples and was 2.22 (95% CI 2.06 to 2.38) for citrate-unadjusted samples. For samples with Ht >62%, the % change between sample pairs was >10%. Results using HS Plus showed a moderate correlation between citrate-adjusted and unadjusted samples (R2 = 0.4267, p < 0.0001). The INR median was 2.51 (95% CI 2.35 to 2.68) for citrate-adjusted samples and 3.45 (95% CI 3.11 to 3.80) for citrate-unadjusted samples. For samples with Ht>55%, the % change between sample pairs was higher than 10%.
Conclusion:
Our data demonstrate that in patients with polycythemia on warfarin therapy, INR-RTF does not require anticoagulant adjustment for assessment of samples with Ht <62%.
Related Concept Videos
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Anticoagulant Drugs: Vitamin K Antagonists and Direct Oral Anticoagulants
Warfarin, a prominent vitamin K antagonist family member, exerts its effect by inhibiting the enzyme VKORC1 (vitamin K epoxide reductase complex 1). By hindering this enzyme, warfarin...
Venous Thrombosis III: Interprofessional Care
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Venous Thrombosis IV: Nursing Management