Video Laryngoscope Intubation With an Aerosol Barrier Device: A Randomized Sequential Crossover Pilot Study
Related Experiment Video
Updated: Dec 2, 2025

Author Spotlight: Advancing Awake Nasotracheal Intubation with Flexible Video Rhino-Laryngoscopes
Published on: August 2, 2024
Masafumi Idei1, Takeshi Nomura1, Philippe Jouvet2
1Department of Intensive Care Medicine, Tokyo Women's Medical University, Tokyo, Japan.
Objectives:
To assess the impact of the use of aerosol barrier device, Splashguard-CG, on the endotracheal intubation with different types of laryngoscope.
Design:
A pilot randomized sequential crossover simulation study.
Setting:
A single academic center in Japan.
Subjects:
Physicians in a single academic university hospital in Japan.
Interventions:
Use of Splashguard-CG.
Measurements And Main Results:
All participants were asked to perform endotracheal intubation to a manikin simulator using three different devices (Macintosh laryngoscope; Airway Scope [Nihon Kohden, Tokyo, Japan]; and McGRATH MAC [Aircraft Medical, Edinburgh, United Kingdom]) with and without Splashguard-CG in place, which required a total of six attempts and measured the intubation time as the primary outcome. Thirty physicians (15 experienced physicians and 15 less-experienced physicians) were included. Intubation time using Macintosh laryngoscope was significantly longer in the group with Macintosh laryngoscope and Splashguard-CG compared with the group without Splashguard-CG by the median difference of 4.3 seconds (interquartile range, 2.6-7.4 s; p < 0.001). There was no significant increase in the intubation time with or without Splashguard-CG for the Airway Scope (0.6 s; interquartile range, -3.7 to 3.2 s; p = 0.97) and the McGRATH MAC (0.5 s; interquartile range, -1.4 to 4.6 s; p = 0.09). This trend was found in both the experienced and less-experienced groups. We observed significant increases of subjective difficulty of the endotracheal intubation evaluated by using a Visual Analog Scale in the Splashguard-CG groups for all three types of devices.
Conclusions:
The use of a video laryngoscope with an aerosol barrier device does not impact the time required endotracheal intubation in a simulation environment. This method can be considered as airway management for coronavirus disease 2019.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Tracheostomy Suctioning II: Procedure
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...