Comparison of PR, QRS, and QT interval measurements by seven ECG interpretation programs
J De Bie1, I Diemberger2, J W Mason3
1Mortara Instrument Europe s.r.l., Bologna, Italy.
Background:
Electrocardiograph-generated measurements of PR, QRS, and QT intervals are generally thought to be more precise than manual measurements on paper records. However, the performance of different programs has not been well compared.
Methods:
Routinely obtained digital electrocardiograms (ECGs), including over 500 pediatric ECGs, were used to create over 2000 10 s analog ECGs that were replayed through seven commercially available electrocardiographs. The measurements for PR interval, QRS duration, and QT interval made by each program were extracted and compared against each other (using the median of the programs after correction for program bias) and the population mean values.
Results:
Small but significant systematic biases were seen between programs. The smallest and largest variation from the population mean differed by 4.7 ms for PR intervals, 5.8 ms for QRS duration, and 12.4 ms for QT intervals. In pairwise comparison programs showed similar accuracy for most ECGs, with the average absolute errors at the 75th percentile for PR intervals being 4-6 ms from the median, QRS duration 4-8 ms, and QT interval 6-10 ms. However, substantial differences were present in the numbers and extent of large, clinically significant errors (e.g at the 98th percentile), for which programs differed by a factor of two for absolute errors, as well as differences in the mix of overestimations and underestimations.
Conclusions:
When reading digital ECGs, users should be aware that small systematic differences exist between programs and that there may be large clinically important errors in difficult cases.
Related Concept Videos
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
ECG Interpretation of Arrhythmias I: Sinus Arrhythmias
Types of Arrhythmias
Sinus Node Arrhythmias
Sinus Bradycardia: Originating from the sinoatrial (SA) node, sinus bradycardia involves slower impulses, resulting in a heart rate of less than 60 beats per minute (bpm). Causes include sleep, vagal stimulation, beta-blockers, hypothyroidism,...