Pulse oximeter plethysmograph waveform and automated oscillometric sphygmomanometer for ankle-brachial index
Cosby G Arnold1, J Richard Walker2, E Jeffrey Metter3
1Department of Emergency Medicine, University of Colorado School of Medicine, Denver, CO, USA.
Objectives:
There are limited non-invasive methods to assess lower extremity arterial injuries in the emergency department (ED) and pre-hospital setting. The ankle-brachial index (ABI) requires careful auscultation by Doppler, an approach made difficult in noisy environments. We sought to determine the agreement of the ABI measured using the pulse oximeter plethysmograph waveform (Pleth) with auscultation by Doppler in a controlled setting. A secondary outcome sought to examine the agreement of ABI by automated oscillometric sphygmomanometer (AOS) with Doppler.
Methods:
We measured blood pressure in the right upper and lower extremities of healthy volunteers using: (1) Doppler and manual sphygmomanometer; (2) Pleth and manual sphygmomanometer; and (3) AOS. The Bland-Altman approach to assessing agreement between methods was used comparing mean differences between ABI pairs to their means for Doppler versus Pleth and Doppler versus AOS. The intraclass correlation coefficient (ICC) from mixed effects models examined intra- and inter-rater reliability.
Results:
Among 100 participants with normal ABI the mean ABI (95%CI) were Doppler 1.11 (0.90-1.33), Pleth 1.10 (0.91-1.30), and AOS 1.10 (0.90-1.30). The ABI difference (95% CI for limits of agreement) were 0.01 (-0.20,0.18) for Doppler-Pleth and 0.02 (-0.26, 0.22) for Doppler-AOS. The ICC for the Doppler-Pleth comparison (ICC = 0.56, 95% CI 0.47-0.63) was greater than for the Doppler-AOS (ICC = 0.32, 95% CI 0.19-0.43).
Conclusions:
The ABI measured using the Pleth has a high level of agreement with measurement by Doppler. The AOS and Doppler have good agreement with greater measurement variability. Pleth and AOS may be reasonable alternatives to Doppler for ABI.
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Pulse
The pulse serves as a clinical...