Digital Variance Angiography in Lower-Limb Angiography with Metal Implants
M B Bastian1, A M König1, S Viniol1
1Department of Diagnostic and Interventional Radiology, University Hospital Marburg, Philipps University of Marburg, Baldingerstrasse 1, 35043, Marburg, Germany.
Digital variance angiography (DVA) significantly improves lower-limb angiography image quality in patients with metal implants. This advanced technique offers better contrast-to-noise ratio (CNR) and visual assessment compared to digital subtraction angiography (DSA).
Area of Science:
- Radiology
- Medical Imaging
- Interventional Cardiology
Background:
- Metal implants in patients can degrade angiographic image quality due to automated beam adjustments.
- Digital subtraction angiography (DSA) is the standard but can be affected by artifacts from implants.
- Digital variance angiography (DVA) has shown potential for superior image quality and contrast-to-noise ratio (CNR).
Purpose of the Study:
- To evaluate if DVA can overcome the image quality reduction caused by metal implants in lower-limb angiographies.
- To compare the effectiveness of DVA versus DSA in the presence of metallic artifacts.
- To assess the impact of DVA on contrast-to-noise ratio (CNR) and overall image quality.
Main Methods:
- Retrospective analysis of 85 lower-limb iodine contrast angiograms from 12 patients with metal implants.
- Processing of angiograms using DVA and comparison with standard DSA.
- Objective measurement of CNR and subjective evaluation of image quality by three experienced radiologists using Likert scales and paired comparisons.
Main Results:
- DVA demonstrated a median CNR ratio (CNRDVA/CNRDSA) of 1.84, indicating significantly higher CNR.
- Paired comparisons showed 84.5% preference for DVA images, with high interrater agreement (83.2%).
- Mean image quality scores were significantly higher for DVA (4.43) compared to DSA (3.64).
Conclusions:
- DVA significantly enhances image quality in lower-limb angiography when metal implants are present.
- The improved CNR with DVA suggests potential for reduced radiation exposure in such procedures.
- DVA represents a promising alternative to DSA for improving imaging in complex patient populations with implants.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
07:53Author Spotlight: Improved Imaging for Neovascular Development in Congenital Heart Disease Research
Published on: April 26, 2024
Related Concept Videos
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Imaging Studies VII: Vascular Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...