Related Experiment Video
Updated: Dec 2, 2025

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Advanced fully integrated radiofrequency/optical-coherence-tomography irrigated catheter for atrial fibrillation
Valentina Tiporlini1, Selam Ahderom1, Peter Pratten1,2
1Electron Science Research Institute, Edith Cowan University, Joondalup, Western Australia, Australia.
Abstract:
The inability of current catheter ablation procedures to accurately monitor lesion formation limits their safety and efficacy. An advanced fully integrated radiofrequency (RF)/optical coherence tomography (OCT) ablation catheter is developed, which enables real-time monitoring during ablation. An OCT fiber array is especially designed, developed and integrated into an off-the-shelf irrigated RF ablation catheter. In-vitro experimental studies performed on poultry and ovine hearts demonstrate the ability of the integrated RF/OCT system to provide information on the quality and orientation of catheter/wall contact. Experimental results show that adipose tissue can be accurately identified from normal myocardial tissue with 94% accuracy and lesion formation is monitored with an overall accuracy of 93%. The ability to predict pop events is also demonstrated, with an accuracy of 86%.
More Related Videos
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization I: Pre-Procedure Overview