Accelerated 3D Left Atrial Late Gadolinium Enhancement in Patients with Atrial Fibrillation at 1.5 T: Technical
Suvai Gunasekaran1, Hassan Haji-Valizadeh1, Daniel C Lee1
1Department of Radiology (S.G., D.C.L., R.J.A., M.M., D.K.) and Department of Internal Medicine, Division of Cardiology (D.C.L., R.S.P.), Northwestern University Feinberg School of Medicine, 737 N Michigan Ave, Suite 1600, Chicago, IL 60611; Department of Biomedical Engineering, Northwestern University, Evanston, Ill (S.G., M.M., D.K.); Department of Medicine, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, Mass (H.H.V.); and Department of Internal Medicine, Division of Cardiovascular Medicine (B.D.W., M.I.), and Department of Radiology and Imaging Sciences (E.G.K.), University of Utah, Salt Lake City, Utah.
Purpose:
To develop an accelerated three-dimensional (3D) late gadolinium enhancement (LGE) pulse sequence using balanced steady-state free precession readout with stack-of-stars k-space sampling and extra motion-state golden-angle radial sparse parallel (XD-GRASP) reconstruction and test the performance for detecting atrial scar and fibrosis in patients with atrial fibrillation (AF).
Materials And Methods:
Twenty-five patients with AF (20 paroxysmal and five persistent; 65 years ± 7 [standard deviation]; 18 men) were imaged at 1.5 T using the proposed LGE sequence with 1.3 mm × 1.3 mm × 2-mm spatial resolution and predictable imaging time. The resulting images were compared with historic images of 25 patients with AF (18 paroxysmal and seven persistent; 67 years ± 10; 14 men) obtained using a reference 3D left atrial (LA) LGE sequence with 1.3 mm × 1.3 mm × 2.5-mm spatial resolution. Two readers visually graded the 3D LGE images (conspicuity, artifact, noise) on a five-point Likert scale (1 = worst, 3 = acceptable, 5 = best), in which the summed visual score (SVS) of 9 or greater was defined as clinically acceptable. Appropriate statistical analyses (Cohen κ coefficient, Mann-Whitney U test, t tests, and intraclass correlation) were performed, where a P value < .05 was considered significant.
Results:
Mean imaging time was significantly shorter (P < .01) for the proposed pulse sequence (5.9 minutes ± 1.3) than for the reference pulse sequence (10.6 minutes ± 2). Median SVS was significantly higher (P < .01) for the proposed (SVS = 11) than reference (SVS = 9.5) 3D LA LGE images. Interrater reproducibility in visual scores was higher for the proposed (κ = 0.78-1) than reference 3D LA LGE (κ = 0.44-0.75). Intrareader repeatability in fibrosis quantification was higher for the reference cohort (intraclass correlation coefficient [ICC] = 0.94) than the prospective cohort (ICC = 0.79).
Conclusion:
The proposed 3D LA LGE method produced clinically acceptable image quality with 1.5 mm × 1.5 mm × 2-mm nominal spatial resolution and 6-minute predictable imaging time for quantification of LA scar and fibrosis in patients with AF. Supplemental material is available for this article. © RSNA, 2020.