Optimized radiological alert thresholds based on device dosimetric information and peak skin dose in vascular
Nicolas Sas1, Florian Magnier2,3, Eléonore Pouget2
1Plateforme régionale de physique médicale, Centre de Lutte Contre le Centre Jean Perrin, 58 Rue Montalembert, 63011, Clermont-Ferrand Cedex 1, France. nicolas.sas@clermont.unicancer.fr.
Objectives:
The National Council on Radiation Protection (NCRP) report no. 168 recommended that during fluoroscopically guided interventions (FGIs), each patient should be monitored when one of the following thresholds is reached: an air kerma > 5 Gy, a kerma area product (KAP) > 500 Gy.cm2, a fluoroscopy time > 60 min, or a peak skin dose (PSD) > 3 Gy. Whereas PSD is the most accurate metric regarding the prevention of radiological risks, it remains the most difficult parameter to assess. We aimed to evaluate the relevance of the other, more accessible metrics and propose new optimized threshold (OT) for improved patient follow-up.
Methods:
Overall, 108 patients who underwent FGI in which at least one NCRP threshold was reached and PSD was measured were considered. The correlation between all metrics was assessed using principal component analysis (PCA). ROC curves and the sensitivity/specificity of both NCRP and OT to predict PSD > 3 Gy were evaluated.
Results:
The PCA shows that FGI can be decomposed with two components based on time and dose variables. Only KAP and kerma were correlated with PSD. The overall sensitivity and specificity of the new OT regarding KAP (67.6/93.0), kerma (97.3/81.7), and time (62.2/62.0) were better compared with NCRP thresholds (97.3/16.9, 40.5/95.4, and 21.6/74.7).
Conclusions:
This study shows that fluoroscopy time is not a relevant metric when used to predict PSDs > 3 Gy. By adapting KAP and kerma thresholds to predict PSD over 3 Gy, patient follow-ups following vascular FGI can be improved.
Key Points:
• In vascular fluoroscopically guided interventions, principal component analysis demonstrates that between fluoroscopy time, KAP, and kerma, only the two last were correlated to the peak skin dose. • Optimized thresholds replacing NRCP ones obtained with ROC curves analysis were 85,451 μGy.cm2, 2938 mGy, and 41 min for KAP, kerma, and fluoroscopy time respectively. • Improvements to trigger patient follow-up after vascular fluoroscopically guided interventions may be obtained by using the optimized thresholds.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Biological Effects of Radiation
Imaging Studies VII: Vascular Imaging