[Comparison of Goldmann tonometry with air-puff tonometry. A study of 159 patients in Abidjan]
A C Kouassi Rebours1, F X Kouassi1, M Soumahoro1
1Service d'ophtalmologie, université Félix-Houphouet-Boigny, CHU de Cocody, Abidjan, Côte d'Ivoire.
Purpose:
The goal of this study was to compare the intraocular pressure (IOP) values obtained with each of the tonometry methods as well as to study the effect of corneal pachymetry on the measurements obtained by each method.
Material And Methods:
We carried out a cross-sectional, descriptive, analytical, prospective study in the ophthalmology department of a multispecialty clinic in Abidjan (Ivory Coast) from April 2018 to August 2018. Patients aged 15 and over were included in the study. Over this period, IOP was measured by Goldman Applanation Tonometer (HaagStreit, AT 900, Switzerland) and air-puff Tonopachymeter (Nidek, NT 530 P, Japan), which also permitted measurement of central optical corneal pachymetry. The data analyzed focused on socio-demographic characteristics, ophthalmologic history, tonometry, and central corneal pachymetry. The comparison of the values obtained by both devices and the effect of central corneal thickness (CCT) on tonometry were analyzed using EpiDataAnalysis V2.2.2.183 software.
Results:
We examined 318 eyes of 159 patients whose mean age was 40±12.57 years, range 16 to 82 years. The gender ratio was 1. A past ophthalmologic history was noted in 79.2% of patients. The mean IOP by Goldmann applanation tonometry was 12.75±3.17mmHg, and that by non-contact air-puff tonometry was 13.83±3.80mmHg. The mean central corneal thickness was 530.50±32.61μm. The difference between IOP measured by non-contact tonometry and that measured by contact tonometry was statistically significant, with P=0.00. There was no statistically significant relationship between central corneal pachymetry and contact tonometry (P=0.80). However, there was a statistically significant relationship between non-contact tonometry and central pachymetry (P=0.00). The values obtained with the air-puff tonometer are more affected by central corneal thickness.
Conclusion:
The values obtained by Goldmann applanation tonometry are not equivalent to those obtained by non-contact air-puff tonopachymetry.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement